BUFFER SOLUTIONS/CAPSULES
It is recommended that you use buffer solutions closest to
your expected measurement range. If you are measuring
approx. 5 pH then calibrate using 7.00 & 4.01 pH buffer
solutions.
Rinse the pH & temperature probes in de-ionised water
before placing in each buffer solution, if this is not
available tap water is acceptable. The buffer solution's
value changes with temperature and so the instrument
automatically compensates for this during calibration.
Therefore only buffer solutions that follow the temperature
coefficients shown in the table below should be used with
these instruments for best accuracy.
temperature
4.01pH
7.00pH
0
4.01
7.12
5
4.01
7.09
10
4.01
7.06
15
4.00
7.04
20
4.00
7.02
25
4.01
7.00
30
4.01
6.99
35
4.02
6.98
40
4.03
6.97
45
4.04
6.97
50
4.06
6.97
55
4.08
6.97
60
4.10
6.98
WARNING: IPA and other solvents may cause damage to the
case and screen of this instrument.
ACCESSORIES
Cleaning Solution
816-040
pH electrode cleaning solution 500 ml
Storage Solution
816-041
pH electrode storage solution - 500 ml
Buffer Solutions
816-050
4.01 pH buffer solution pink - 100 ml
816-051
7.00 pH buffer solution yellow - 100 ml
816-052
10.01 pH buffer solution blue - 100 ml
Buffer Capsules
816-004
4 pH buffer capsules - pack of 10
816-007
7 pH buffer capsules - pack of 10
816-009
9 pH buffer capsules - pack of 10
816-010
10 pH buffer capsules - pack of 10
pH Electrodes
10.01pH
823-504
Budget pH electrode - 0 to 60 °C
10.32
823-501
General purpose pH electrode - 0 to 100 °C
10.25
823-502
12 mm Spear Combination pH electrode
10.18
823-503
6 mm Spear Combination pH electrode
823-514
Knife probe electrode
10.12
10.06
10.01
9.97
9.93
9.89
9.86
9.83
9.81
9.79
Electronic Temperature Instruments Ltd
Worthing · West Sussex · BN14 8HQ
01903 202151 · sales@etiltd.com · etiltd.com
Manufactured by
545-080/19.01.21
8000/8100
pH METERS
Operating Instructions
Need help?
Do you have a question about the 8000 and is the answer not in the manual?
Questions and answers