C Series Conductivity Sensor Operating Manual
®
Wiring
4-20mA 2-wire
Blue: mA-
Brown: mA+
Typical Installation
Maintenance of conductive electrode
Cleaning of conductive electrode
1. The organic component contamination on the electrode can be cleaned with warm water containing detergent, or it can be cleaned
with alcohol.
2. Calcium and magnesium precipitates are best cleaned with 10% citric acid.
3. The bright platinum electrode can be mechanically cleaned with a soft brush. Ensure that no nicks or scratches are made on the
electrode sensor.
4. The platinum electrode plated with platinum black should only be cleaned using the chemical method. The coating (platinum black) on
the electrode surface will be damaged if cleaning with a soft brush.
Warranty
Icon Process Controls warrants this product to be free from significant deviations in material and workmanship for a period of one year
from the date of purchase. If repair is necessary and has not been the result of abuse or misuse within the warranty period, please return to
Icon Process Controls and amendment will be made without any charge. Icon Process Controls Customer Service Center will determine if
product problem is due to deviations or customer abuse. Out of warranty products will be repaired on a charge basis.
Authorization must be obtained from Icon Process Controls Customer Service Center to issue a RIR number before returning items for
any reason. When applying for authorization, please include date requiring the reason of return. Instruments must be carefully packed to
prevent damage in shipment and insured against possible damage or loss. Icon Process Controls will not be responsible for any damage
resulting from careless or insufficient packing.
Warning:
Damage as a result of inadequate packaging is the User / distributor's responsibility. Please follow the guidelines below before
transporting.
Use the original packaging material if possible, when transporting back the unit for repair. Otherwise wrap it with bubble pack and use a
corrugated box for better protection. Include a brief description of any faults suspected for the convenience of Customer Service Center, if
possible. If there are any questions, feel free to contact our Customer Service Center or distributors.
Phone: 905.469.9283 • Sales: sales@iconprocon.com • Support: support@iconprocon.com
4-20mA 4-wire
Transparent: 4-20mA
Black (thick): Ref
Red: Temperature
Black: Temperature
Connects directly to
ProCon
controller
®
Output
Intake
Do Not use a screwdriver or sharp object
to clean the electrode surface
by
4-20mA + RS485
Output
Red: 9-24VDC +
Thicker
Wire
Black: 9-24VDC -
Transparent: 4-20mA
Black (thick): Ref
Green: RS 485 A
White : RS 485 B
3