Technical specifications
8.1 General technical specifications
Input
0 ... 200 psi g
0 ... 300 psi g
0 ... 500 psi g
0 ... 750 psi g
0 ... 1 000 psi g
Measuring range for absolute
pressure
(only for U.S. market)
0 ... 10 psi a
0 ... 15 psi a
0 ... 20 psi a
0 ... 30 psi a
0 ... 60 psi a
0 ... 100 psi a
0 ... 150 psi a
0 ... 200 psi a
0 ... 300 psi a
Output
Current signal
Voltage signal
Characteristic curve
Measuring accuracy
Measurement deviation at 25 °C (77 °F), including
characteristic curve deviation, hysteresis and re‐
peatability
Set-up time T99
Long-term drift
Effect of ambient temperature
26
Load
Auxiliary power U
B
Current consumption I
B
Load
Auxiliary power U
B
Current consumption
Start of scale value and measuring span
Start of scale value and measuring span
Influence of vibration (per IEC 60068-2-6)
Effect of auxiliary power supply
≥ -14.5 / ≤ 550 psi g
≥ -14.5 / ≤ 800 psi g
≥ -14.5 / ≤ 1400 psi g
≥ -14.5 / ≤ 2000 psi g
≥ -14.5 / ≤ 2000 psi g
Overload limit
≥ 0 / ≤ 35 psi a
≥ 0 / ≤ 35 psi a
≥ 0 / ≤ 50 psi a
≥ 0 / ≤ 80 psi a
≥ 0 / ≤ 140 psi a
≥ 0 / ≤ 200 psi a
≥ 0 / ≤ 350 psi a
≥ 0 / ≤ 550 psi a
≥ 0 / ≤ 800 psi a
4 ... 20 mA
(U
- 10 V) / 0.02 A
B
7 ... 33 V DC
≤ 20 mA
0 ... 10 V DC
≥ 10 kΩ
12 ... 33 V DC
< 7 mA at 10 kΩ
Linear rising
Typical: 0.25% of full-scale value
Maximum: 0.5% of full-scale value
< 5 ms
0.25% of full-scale value/year
0.25%/10 K of full-scale value
0.005%/g to 500 Hz in all directions
0.005%/V
Compact Operating Instructions, 09/2020, A5E03304326-AA
> 550 psi
> 800 psi
> 1400 psi
> 2000 psi
> 2000 psi
Burst pressure
> 35 psi
> 35 psi
> 50 psi
> 80 psi
> 140 psi
> 200 psi
> 350 psi
> 550 psi
> 800 psi
(10 ... 30 V for Ex)