Table 9-5 - Digital command and control interface specifications (Continued)
Commands
Data outputs
Table 9-6 - Power specifications
Supply voltage
Power consumption
Protection
Table 9-7 - Environmental specifications
Operating temperature
Storage temperature
Shock
Magnetic
Ingress protection
16889R10 • 2020-11-12
XYZ/UVW mode switching
l
Calibration channel selection (off, enable all, U, V, or W)
l
Short period/long period mode switching
l
Firmware updates
l
State-of-health display
l
Upload user calibration data
l
Independent mass position values
l
Instrument temperature
l
Factory sensitivity and sensor response data
l
User calibration data (poles and zeros)
l
Instrument serial number and firmware revision
l
9 V to 36 V DC isolated inputs
See
"Power Consumption" on page 56
consumption scenarios
Reverse-voltage and over-voltage protected
l
Self-resetting over-current protection
l
For models TC120-SV1, TC20-SV1, TC120-PH2 and TC20-
PH2: -20ºC to 60ºC (Standard Model)
For models TC120-PH2-XC and TC20-PH2-XC: -50ºC to 60ºC
(Polar Certified Model)
-40ºC to 70ºC
100 g half sine, 5 ms without damage, 6 axes
l
No mass lock required for transport
l
Insensitive to natural variations of the Earth's magnetic field
For models TC120-SV1 and TC20-SV1: Rated to IP67 for
outdoor use, dust, and immersion resistance
For models TC120-PH2, TC20-PH2, TC120-PH2-XC and
TC20-PH2-XC: Rated to IP68 to 300 m for prolonged
immersion