1 About this manual 1.1 Validity This operating manual is for customers of Pfeiffer Vacuum. It describes the func- tioning of the designated product and provides the most important information for safe use of the unit. The description follows applicable EU guidelines. All informa- tion provided in this operating manual refer to the current state of the product's de- velopment.
Safety Work instruction: here you have to do something. Instructions in the text Abbreviations used ONF: Oil mist filter ORF: Oil return device 2 Safety 2.1 Safety precautions NOTE Duty to inform Each person involved in the installation, operation or maintenance of the vacuum pump must read and observe the safety-related parts of these operating instructions.
4 Product description 4.1 Product identification To correctly identify the product when communicating with Pfeiffer Vacuum, al- ways have the information from the rating plate available. • Model and model number •...
Operating fluid drain screw 30 10 Fig. 2: ONF 16/25/25 L Operating fluid return If the operating fluid accumulated in the ONF reaches a specified level, the operat- ing fluid is channeled via differential pressure back into the rotary vane pump. Con-...
Installation 5 Installation 5.1 Assembly To install the ONF in a vacuum system, an ISO small flange or ISO clamp flanges are provided on the input side and output side. The flanges are provided with pro- tective caps when delivered to protect the seals. If gases are pumped that may not enter the atmosphere, an exhaust line must be connected to the exhaust flange.
5.2 Installing the operating fluid return line Unscrew operating fluid drain screw 33. ONF 16/25/25 L Drain off operating fluid if so and fill in the pump. Screw in hose nipple 241 inplace of the operating fluid drain screw; take care with O-ring.
Operation (241/145) (78) Fig. 4: PentaLine pump with ONF and Operating fluid return line Operating fluid drain screw 145 O-ring 242 Operating fluid return line Screw fitting 241 Hose nipple 243 Hose clip Sealing plug 6 Operation NOTE Make sure that the return for the operating fluid works. Operating fluid is only sucked in and returned from an operating pressure of <...
Maintenance 7 Maintenance 7.1 Draining the operating fluid If the accumulated operating fluid in the ONF is above the top edge of the sight- glass 12, the operating fluid must be drained. The intervals at which the operating fluid is drained depends on the operating con- ditions.
The filter insert must be exchanged or cleaned when the exhaust pressure increas- es until the pressure relief valve opens inside the ONF and oil mist exits. Dismantling 20 16 Fig. 5: ONF 16/25/25 L (with ONF 25 XL 8.1 is a part of 8) Casing Sight glass Compression spring Cover...
Page 12
Maintenance Check the sealed surface of valve buffer 14; replace if necessary. – When assembling lightly oil valve buffer and ensure the correct seating in the valve housing 6. Assembling Continue assembly by reversing the dismantling sequence, paying particular at- tention to compliance with the following steps.
Fill out the "Service Request" form and send it by fax or e-mail to your Pfeiffer Vacuum service address. Include the confirmation on the service request from Pfeiffer Vacuum with your shipment. Fill out the declaration on contamination and include it in the shipment (re- quired!).
Technical data 10.1 Dimensions Fig. 6: ONF 16/25/25 L Dimensions ONF 16 ONF 25 ONF 25 L 142 mm 147 mm 218 mm 80 mm 80 mm 80 mm 120 mm 132 mm 134 mm 127 mm 127 mm 182 mm...
The spare parts packages listed here are only applicable for standard models. Please state all information on the rating plate when ordering spare parts. Other spare parts than those described in this manual must not be used without the agreement of Pfeiffer Vacuum. 11.1 Spare parts packages Spare parts pack-...
Disposal ONF 025 XL Fig. 9: ONF 025 XL 12 Disposal Products or parts thereof (mechanical and electrical components, operating fluids, etc.) may cause environmental burden. Safely dispose of the materials according to the locally applicable regulations.
Need help?
Do you have a question about the ONF 16 and is the answer not in the manual?
Questions and answers