PCB Piezotronics Inc. claims proprietary rights in
the information disclosed hereon. Neither it nor any
reproduction thereof will be disclosed to others
without the written consent of PCB Piezotronics Inc.
B
A
1
FASTEN CABLE TO TEST STRUCTURE TYPICALLY WITHIN 2-3"[51-76] OF SENSOR, THEN FASTEN AGAIN WITHIN
3-4"[76-101] OF PREVIOUS ATTACHMENT. BETWEEN THE TEST STRUCTURE AND A FIXED STRUCTURE, ALLOW A SERVICE
LOOP LARGE ENOUGH TO PREVENT PULLING OF THE CABLE WHEN SHAKING. MORE ATTACHMENT POINTS WILL
PROVIDE LESS NOISE IN THE RESULTING DATA. LOOSE CABLES OR PARTS ELSEWHERE ON THE TEST STRUCTURE CAN
ALSO GENERATE "NOISE" ON THE SIGNAL RECEIVED FROM THE MODEL 3713 SERIES.
4
3
MODEL 037EN
9-PIN EN PLUG TO
9-PIN EN PLUG
3-4[76-101]
2-3[51-76] 1
3
2
TAPE, PETRO WAX,
OR EQUIVALENT (TYPICAL)
1
TEST STRUCTURE
UNLESS OTHERWISE SPECIFIED TOLERANCES ARE:
DIMENSIONS IN INCHES
DIMENSIONS IN MILLIMETERS
[ IN BRACKETS ]
DECIMALS
XX ±.03
DECIMALS
X ± 0.8
XXX ±.010
XX ± 0.25
ANGLES
2 DEGREES
ANGLES
2 DEGREES
FILLETS AND RADII
FILLETS AND RADII
.003 - .005
0.07 - 0.13
2
1
REVISIONS
REV
DESCRIPTION
-SEE SHEET 1
"SERVICE LOOP" TO ALLOW MOTION
WITHOUT PULLING CABLE
DRAWN
CHECKED
ENGINEER
JES
3/20/19
JDM
3/20/19
NF
3/20/19
3425 WALDEN AVE. DEPEW, NY 14043
TITLE
(716) 684-0001 E-MAIL: sales@pcb.com
OUTLINE DRAWING
DWG. NO.
MODEL 3713F11XXXG
CODE
IDENT. NO.
TRIAXIAL ACCELEROMETER
52681
SCALE:
1.5X
1
DIN
B
A
70537
SHEET
2 OF 2