Gaumard NOELLE S554.100 User Manual page 69

Maternal birthing simulator
Table of Contents

Advertisement

Streaming audio does not work, tab is not displayed
The backup battery on the
simulator is depleted
UNI has set the power mode to STAND-BY automatically
Wired communication module
is not connected
"Wired communication module not found" message is displayed when UNI is started
Wired communication is
module not identified by the
computer
Drivers not properly loaded.
Is the communication indicator
panel consistantly yellow?
CPR is not detected
Compression sensor is not
calibrated
Ventilations are not detected
Noelle's delivery baby is disengaged prematurely
Wrong force sensor reading.
Plug charger into the simulator. Verify LED light on charger indicates
"charging" status.
NOELLE should always be plugged in while being operated. The birth
mechanism is power intensive and will drain the backup battery quickly.
Reconnect Newborn to the charger.
Connect the communication module to any USB port.
Close the software and try disconnecting the communication module for at
least five seconds, then plug it back in and restart the software
Perform a full shutdown procedure on the tablet.
See solution above in section making reference to "Communication with the
simulator cannot be established"
See Calibration section inside the User Guide.
See Calibration section inside the User Guide.
Set the Respiratory Rate to 0.
When baby is pulled with more than 35 lbs. of force, to avoid damage, the
baby is released. If user is not pulling the baby it means that "Force Sensor" is
offset. See the "Calibration Wizard" section inside User's Manual and reset the
force sensor to its default value ("Reset to Default Force" button).
NOELLE S554.100 |
User Manual
Other | 58