Finding Heating Tubes In The Floor; 11�10 Locating A Complete Electrical Installation - VOLTCRAFT LSG-10 Operating Instructions Manual

Cable detector
Table of Contents

Advertisement

Available languages

Available languages

11.9 Finding heating tubes in the floor

For the best search results, the heating pipes should
be disconnected from the earth connection� Connect
the transmitter to the heater's metal pipe and an earth
connection�
Note:
– Suitable earth connections are any power
outlet protective conductors�
Trace the pipe's course with the receiver�
11.10 Locating a complete electrical installation
Switch off the fuses and power down the
whole system� This measurement requires
manipulating the main electrical distribution
system and must only be performed by an
electrician�
With the cable detector, all present outlets and cables
can be determined and checked for contact� Remove
the main distribution bridge between the protective con-
ductor "PE" and the neutral conductor "N"�
Connect the transmitter to the distribution main of "N"
and "PE"�
The neutral conductor can be traced through the com-
plete system with the receiver�
43

Advertisement

Table of Contents
loading

This manual is also suitable for:

2633262

Table of Contents