PTW MP3-PL Assembly Instructions And User's Manual

PTW MP3-PL Assembly Instructions And User's Manual

Therapy beam analyzer
Table of Contents

Advertisement

Assembly Instructions and User Manual
Therapy Beam Analyzer T41050
Operating Manual
Technical Manual
Service Manual
MP3-PL
Version 0 or higher
Contents
D850.131.05/01 en
2016-05
Bru

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the MP3-PL and is the answer not in the manual?

Questions and answers

Summary of Contents for PTW MP3-PL

  • Page 1 Assembly Instructions and User Manual MP3-PL Therapy Beam Analyzer T41050 Version 0 or higher Contents  Operating Manual  Technical Manual  Service Manual D850.131.05/01 en 2016-05...
  • Page 2: General Information

    The product is a class IIb device by PTW-Freiburg or by persons authorized by (MDD). PTW-Freiburg, and if the product is used in compliance with the technical documentation.  Electrical Safety ...
  • Page 3 MP3-PL Therapy Beam Analyzer T41050 General Information PTW-FREIBURG Physikalisch-Technische Werkstätten Dr. Pychlau GmbH Lörracher Str. 7 79115 FREIBURG GERMANY Phone: +49 761 49055-0 Fax: +49 761 49055-70 info@ptw.de www.ptw.de D850.131.05/01 en...
  • Page 4: Table Of Contents

    Contents General Information Contents Illustrations Intended Use Safety Information Operating Manual The Components of the MP3-PL System Perspex Tank Moving Mechanism Positioning Device SCANLIFT Carriage and Reservoir Control Unit and Control Pendant Assembly of the System Assembling SCANLIFT, Positioning Device, Chamber Cable Holding Device and Perspex Tank...
  • Page 5 MP3-PL Therapy Beam Analyzer T41050 Contents Service Manual Cleaning Preventive Maintenance 10 Storage 11 Disposal of the Product Literature Appendix: China Electronic Industry Standard Compliance Index D850.131.05/01 en...
  • Page 6: Illustrations

    Figure 3: MP3-PL System with accessories for measurements with a measuring and a reference detector Figure 4: MP3-PL Perspex tank with junction box and moving mechanism (top: front view; below: rear view) Figure 5: Connections at the junction box Figure 6:...
  • Page 7: Intended Use

    MULTIDOS, the TBA CONTROL UNIT, the TBA synchrotron). CONTROL PENDANT, the PC and the ionization chambers. For filling or emptying the MP3-PL Therapy Beam Analyzers the SCANLIFT carriage and reservoir is used. NOTE The basic system consists of the following compo-...
  • Page 8: Safety Information

    MP3-PL Therapy Beam Analyzer T41050 Safety Information Safety Information This is the safety alert symbol. It is used to Safety Information alert the user to potential hazards. Obey all DANGER safety messages that follow this symbol to avoid possible bodily injury or equipment Operation in areas where an explosion hazard damage.
  • Page 9 MP3-PL Therapy Beam Analyzer T41050 Safety Information WARNING WARNING  Liquids must not enter the product. If liquids Improper handling. have entered the product, it must be thor- Patient Hazard! oughly inspected before being used again. The product is a medical electrical device and ...
  • Page 10 MP3-PL Therapy Beam Analyzer T41050 Safety Information WARNING WARNING Electricity is a source of risk, particularly when Magnetic and electrical fields are capable of in- the product is not in perfect operating condition terfering with the proper performance of the de- or when it is operated inappropriately.
  • Page 11 Use the product only in conjunction with the shifted (e.g., laterally by 15 mm). Continued irra- products approved for this purpose by PTW diation through the original position can lead to and/or the products listed in the section "Intend- radiation damage of the entrance window mate- ed Use".
  • Page 12 If the water phantom is lowered to the bottom When used with the SCANLIFT, rotating the position by means of the SCANLIFT lift mecha- MP3-PL by means of the positioning device is nism, only very little space (approx. 10 –12 mm) not allowed.
  • Page 13: Figure 1: Definition Of Patient Environment

    MP3-PL Therapy Beam Analyzer T41050 Safety Information Symbols on the Nameplate NOTE Set up the device so that the operator has a Symbol Description clear, unobstructed view of the indicators and controls and can operate the controls easily. Please refer to user manual! NOTE The product bears the CE-mark.
  • Page 14 MP3-PL Therapy Beam Analyzer T41050 Safety Information D850.131.05/01 en...
  • Page 15: Operating Manual

    MP3-PL Therapy Beam Analyzer T41050 Operating Manual – The Components of the MP3-PL System Operating Manual 1 The Components of the MP3-PL System Figure 2: MP3-PL System with accessories for measurements with 24 PinPoint chambers D850.131.05/01 en...
  • Page 16: Figure 3: Mp3-Pl System With Accessories For Measurements With A Measuring And A Reference Detector

    MP3-PL Therapy Beam Analyzer T41050 Operating Manual – The Components of the MP3-PL System Figure 3: MP3-PL System with accessories for measurements with a measuring and a reference detector D850.131.05/01 en...
  • Page 17: Perspex Tank

    MP3-PL Therapy Beam Analyzer T41050 Operating Manual – The Components of the MP3-PL System Perspex Tank The moving mechanism as well as several ancillary devices (refer to section 2) are attached to the Per- spex tank. The tank (Figure 4) is filled with water. The water level should remain approx.
  • Page 18: Moving Mechanism

    MP3-PL Therapy Beam Analyzer T41050 Operating Manual – The Components of the MP3-PL System Moving Mechanism CAUTION Improper handling. Risk of injury! Do not reach into the Perspex tank while the moving mechanism is working. Pay attention that hairs or cloths of the user and other persons involved do not touch the rotating spindle.
  • Page 19: Positioning Device

    MP3-PL Therapy Beam Analyzer T41050 Operating Manual – The Components of the MP3-PL System Positioning Device The positioning device allows the Perspex tank to be aligned, using the laser system of the accelera- tor and the lines on the walls of the tank.
  • Page 20: Scanlift Carriage And Reservoir

    MP3-PL Therapy Beam Analyzer T41050 Operating Manual – The Components of the MP3-PL System SCANLIFT Carriage and Control Unit and Control Reservoir Pendant The MP3-PL system is operated via the control unit CAUTION and the control pendant. Improper handling. Equipment damage!
  • Page 21: Assembly Of The System

    MP3-PL Therapy Beam Analyzer T41050 Operating Manual – Assembly of the System 2 Assembly of the System  Place the Perspex tank (6) on the base plate of NOTE the chamber cable holding device. Pay attention You can use the supplied tools to assemble the that the water connection of the Perspex tank (7) system.
  • Page 22: Figure 7: Assembling Scanlift, Positioning Device, Chamber Cable Holding Device And Perspex Tank

    MP3-PL Therapy Beam Analyzer T41050 Operating Manual – Assembly of the System Figure 7: Assembling SCANLIFT, positioning device, chamber cable holding device and Perspex tank D850.131.05/01 en...
  • Page 23: Assembling The Junction Box And The Moving Mechanism

    MP3-PL Therapy Beam Analyzer T41050 Operating Manual – Assembly of the System Assembling the Junction 1 Bracket of the chamber cable holding device Box and the Moving 2 Screw M5x10 (16x) Mechanism 3 Base plate of the chamber cable holding device...
  • Page 24: Figure 8: Mounting The A-Axis

    MP3-PL Therapy Beam Analyzer T41050 Operating Manual – Assembly of the System  Check that the two bevel gear wheels engage properly in the centering shoulder and adjust, if necessary, by turning the threaded control shaft of the C-axis. ...
  • Page 25: Figure 9: Mounting The B-Axis

    MP3-PL Therapy Beam Analyzer T41050 Operating Manual – Assembly of the System Figure 9: Mounting the B-axis 1 Gear box B 8 Screw M3x12 (2x) 2 Screw M5x16 (2x) 9 Cable holder (2x) 3 Conical bolt 10 Motor cable B/C...
  • Page 26: Figure 10: Mounting The C-Axis

    MP3-PL Therapy Beam Analyzer T41050 Operating Manual – Assembly of the System Figure 10: Mounting the C-axis 1 Catch pin 6 Slider B 2 Groove 7 Centering bore 3 Slider C 8 Centering shoulder 4 Axis C 9 Screw M4x10 (4x)
  • Page 27: Installation Of The System For Measurement

    MP3-PL Therapy Beam Analyzer T41050 Operating Manual – Installation of the System for Measurement 3 Installation of the System for Measurement  Align the Perspex tank by using the laser system NOTE of the accelerator. You can use the supplied tools to install the sys- tem.
  • Page 28: Mounting Detectors On The Moving Mechanism

    MP3-PL Therapy Beam Analyzer T41050 Operating Manual – Installation of the System for Measurement  Mounting Detectors on the Place the holder T21003.1.020 on a flat base. Alternatively mount holder Moving Mechanism T21003.1.020 on the robot’s table top using the two M6x35 screws and the washers.
  • Page 29: Figure 12: Vertical Mounting Of The 3D-Detectorblock With 24 Pinpoint Chambers

    MP3-PL Therapy Beam Analyzer T41050 Operating Manual – Installation of the System for Measurement Figure 13: Horizontal mounting of the 3D-detector- block with 24 PinPoint chambers 1 3D-detectorblock T21003.1.001 Figure 12: Vertical mounting of the 3D-detector- 2 Holder T21003.1.010 block with 24 PinPoint chambers 3 Clamping screw (2x) 1 3D-detectorblock T21003.1.001...
  • Page 30: Figure 14: Inserting The 24 Pinpoint Chambers Into The Linear Detector Holder

    MP3-PL Therapy Beam Analyzer T41050 Operating Manual – Installation of the System for Measurement 3.2.2 Linear Detector Holder with 24 PinPoint Chambers Type 31015  Place the holder T21004.1.003 on a flat base. Alternatively mount holder T21004.1.003 on the robot’s table top using the two M6x30 screws and the washers.
  • Page 31: Figure 15: Vertical Mounting Of The Linear Detector Holder With 24 Pinpoint Chambers

    MP3-PL Therapy Beam Analyzer T41050 Operating Manual – Installation of the System for Measurement Vertical mounting refer to Figure 15  Mount the holder T21004.1.004 on the C-slider with the two M4x16 screws (4, Figure 15).  Mount the linear detector holder with the Pin- Point chambers on the holder T21004.1.004 us-...
  • Page 32: Figure 16: Horizontal Mounting Of The Linear Detector Holder With 24 Pinpoint Chambers

    MP3-PL Therapy Beam Analyzer T41050 Operating Manual – Installation of the System for Measurement Horizontal mounting refer to Figure 16  Mount the linear detector holder with the Pin- Point chambers on the holder T21004.1.004 us- ing the two M4x20 screws (4, Figure 16).
  • Page 33: Figure 17: Horizontal Mounting Of The Bragg Peak Chamber

    MP3-PL Therapy Beam Analyzer T41050 Operating Manual – Installation of the System for Measurement 3.2.3 Bragg Peak Chamber Type 34073 and Use following holders for the Bragg Peak cham- Bragg Peak Chamber Type 34070 bers:  Bragg Peak chamber Type 34073 with holder CAUTION T41029.1.200 and...
  • Page 34: Figure 18: Vertical Mounting Of The Bragg Peak Chamber

    MP3-PL Therapy Beam Analyzer T41050 Operating Manual – Installation of the System for Measurement Figure 18: Vertical mounting of the Bragg Peak chamber 1 Bragg Peak chamber 2 Holder 3 Clamping screw (2x) 4 C-slider 5 C-axis 6 Screw M4x10 (2x)
  • Page 35: Connecting The Detectors

    MP3-PL Therapy Beam Analyzer T41050 Operating Manual – Installation of the System for Measurement  Connecting the Detectors Bundle the chamber cables with the supplied Velcro strips as shown in Figure 19. 3.3.1 24 PinPoint Chambers Type 31015 with ...
  • Page 36: Figure 20: Chamber Cable Holding Device T20015.1.200

    MP3-PL Therapy Beam Analyzer T41050 Operating Manual – Installation of the System for Measurement Positioning of the Chamber Cable Holding De- vice for Measurements: CAUTION Improper handling. Risk of injury! The chamber cable holding device is very heavy (several kilograms).
  • Page 37: Mounting And Connecting The Reference Detector

    MP3-PL Therapy Beam Analyzer T41050 Operating Manual – Installation of the System for Measurement Mounting and Connecting 3.3.2 Bragg Peak Chamber Type 34073 the Reference Detector  Attach the plug holder T4316/U161 to the Per- spex tank.  If required, connect the ionization chamber to the extension cable T2954/K2-6.
  • Page 38: Figure 22: Mounting The Reference Detector Holder

    MP3-PL Therapy Beam Analyzer T41050 Operating Manual – Installation of the System for Measurement   Insert the reference detector into the holder as Connect the reference detector to detector input shown in Figure 21 as far as it will go. Make labeled CHANNEL 2 on the rear side of the sure the side labeled "Focus"...
  • Page 39: Mounting The Inhomogeneous Phantom T43031

    MP3-PL Therapy Beam Analyzer T41050 Operating Manual – Installation of the System for Measurement Mounting the Inhomo- Mounting the Hemisphere geneous Phantom T43031 T41029.3.231 Figure 23: Mounting the inhomogeneous phantom Figure 24: Mounting the hemisphere T41029.3.231 T43031 1 Screw M6x16 (4x) 1 Screw M6x16 (4x) 2 Holder T41029.1.230...
  • Page 40: Mounting The Wedge T41029.1.220

    MP3-PL Therapy Beam Analyzer T41050 Operating Manual – Installation of the System for Measurement Mounting the Wedge Mounting the Film Holding T41029.1.220 Device T40040 Figure 25: Mounting the wedge T41029.1.220 Figure 26: Mounting film holding device T40040 1 Screw M6x16 (4x) 1 Screw M6x20 (4x) 2 Wedge T41029.1.220 with holder...
  • Page 41: Connecting Scanlift

    MP3-PL Therapy Beam Analyzer T41050 Operating Manual – Installation of the System for Measurement Connecting SCANLIFT 3.9.3 Disconnecting SCANLIFT from Perspex Tank 3.9.1 Preparing SCANLIFT CAUTION refer to Figure 27 Improper handling.  Equipment damage! Put the hose clip onto the open tubing end of the water hose on the top of the SCANLIFT.
  • Page 42: Filling With Water

    Perspex tank with water up to the top horizontal energy radiation. line, the pumping process has to be started again and stopped manually. PTW recommends the following procedure for initial filling of the system: Pumping from SCANLIFT Water Reservoir into ...
  • Page 43 MP3-PL Therapy Beam Analyzer T41050 Operating Manual – Installation of the System for Measurement  Pumping from the Perspex Tank into SCANLIFT First empty the Perspex tank by draining the Water Reservoir water from the Perspex tank into the SCANLIFT water reservoir.
  • Page 44: Connecting The Control Unit And Control Pendant

    MP3-PL Therapy Beam Analyzer T41050 Operating Manual – Installation of the System for Measurement 3.11 Connecting the Control Unit Front Panel and Control Pendant CONTROL PENDANT socket: connection for TBA CONTROL PENDANT CAUTION Direct or scattered radiation NOTE Equipment Damage!
  • Page 45: Installation Check

    MP3-PL Therapy Beam Analyzer T41050 Operating Manual – Installation Check 4 Installation Check Before starting the MP3-PL system, please check When the devices, e.g. TBA CONTROL UNIT and the following points carefully: MULTIDOS, have been switched on, you will get the following indications: ...
  • Page 46: Error Messages And Troubleshooting

    MP3-PL Therapy Beam Analyzer T41050 Operating Manual – Error Messages and Troubleshooting 5 Error Messages and Troubleshooting If problems occur, please check the installation as NOTE outlined in section 2. The following list of possible When contacting our Service team, please give problems may be a useful guideline for trouble- the full serial number of your device.
  • Page 47: Technical Manual

    MP3-PL Therapy Beam Analyzer T41050 Technical Manual – Technical Specifications Technical Manual 6 Technical Specifications Only values with specified tolerances or limits are guaranteed. Values without tolerances are for information purposes only. Description MP3-PL Therapy Beam Analyzer, T41050 Manufacturer PTW-Freiburg...
  • Page 48 MP3-PL Therapy Beam Analyzer T41050 Technical Manual – Technical Specifications Moving mechanism Moving range A-axis 500 mm Moving range B-axis 408.5 mm Moving range C-axis 500 mm Reproducibility 0.1 mm Positioning accuracy 0.5 mm Step interval 0.1 mm Max. cable length...
  • Page 49: Order Information

    TBA control cable T43164/K1-4 TBA motor cable T43164/K2-4 SCANLIFT (incl. control unit) T41024 MP3-M positioning device, swiveling T41022.1.021 Chamber cable holding device for MP3-PL T20015.1.200 Detector connection box 12 x M T16007 PinPoint chamber 0.03 cm³ TM31015 3D-detectorblock T21003 MULTIDOS...
  • Page 50 MP3-PL Therapy Beam Analyzer T41050 Technical Manual – Order Information Accessories 3D detector block T21003 Linear detector holder T21004 Bragg peak chamber 10.5 cm³ TM34070-2,5 Bragg peak chamber 2.5 cm³ TM34073-1,08 Thin-window Bragg peak chamber 10.5 cm³ TM34080-2,5 Chamber holder for type 34070 / type 34080 T41029.1.250...
  • Page 51: Service Manual

    WARNING In addition, ensure that all other service tasks are Liquids have entered the product. only performed by PTW-Freiburg or by a person Shock Hazard, Erroneous Measurement! authorized by PTW-Freiburg. The device must be thoroughly inspected by service personnel before being used again.
  • Page 52 Service Manual – Cleaning Device Surfaces Perspex Tank  First of all, disconnect the MP3-PL system from The Perspex tank is drained by pumping the water the power line, before attempting to clean the into the SCANLIFT water reservoir. device surfaces.
  • Page 53: Preventive Maintenance

    It is recommend- ed to perform a technical inspection of the measur- Check before each use ing system every 2 years at PTW-Freiburg or a qualified calibration laboratory.  Before each use, visually inspect the device and the cables for signs of mechanical damage.
  • Page 54: Storage

    10 Storage 11 Disposal of the Product To preserve a constant evenness of the plates The typical lifetime of the MP3-PL Therapy Beam made of PMMA (e.g. holder for inhomogeneous Analyzer system is about 10 years. phantom), pay attention that these plates are stored...
  • Page 55: Literature

    MP3-PL Therapy Beam Analyzer T41050 Literature Literature Council Directive 93/42/EEC concerning medical devices (Medical Device Directive - MDD) Council Directive 97/43/EURATOM on health protection of individuals against the dangers of ionizing radiation in rela- tion to medical exposure IEC 61010-1 Safety requirements for electrical equipment for measure- ment, control and laboratory use;...
  • Page 56: Appendix: China Electronic Industry Standard Compliance

    MP3-PL Therapy Beam Analyzer T41050 Appendix: China Electronic Industry Standard Compliance Appendix: China Electronic Industry Standard Compliance 附录:中华人民共和国电子产业标准 This Supplement concerns China Electronic Industry Standard Compliance 此附录涉及中华人民共和国电子产业标准的相关规定 The following product pollution control information is provided according to SJ/T11364-2006 Marking for Control of Pollution caused by Electronic Information Products.
  • Page 57  Data listed in the table represents best information available at the time of publication.  This table shows where these substances may be found in the supply chain of PTW electronic information products, as of the date of sale of the enclosed product. Note that some of the component types listed above may or may not be a part of the enclosed product.
  • Page 58: Index

    MP3-PL Therapy Beam Analyzer T41050 Index Index Linear detector holder..........30 Literature ..............55 3D-detectorblock ............28 Maintenance ............... 53 Moving mechanism ............. 18 A-axis, mounting ............23 Moving range limits ............. 18 Adjust Perspex tank ............ 19 Align the tank .............. 17 Anti-corrosion fluid ............