Assembly Instructions and User Manual MP3-PL Therapy Beam Analyzer T41050 Version 0 or higher Contents Operating Manual Technical Manual Service Manual D850.131.05/01 en 2016-05...
The product is a class IIb device by PTW-Freiburg or by persons authorized by (MDD). PTW-Freiburg, and if the product is used in compliance with the technical documentation. Electrical Safety ...
Page 3
MP3-PL Therapy Beam Analyzer T41050 General Information PTW-FREIBURG Physikalisch-Technische Werkstätten Dr. Pychlau GmbH Lörracher Str. 7 79115 FREIBURG GERMANY Phone: +49 761 49055-0 Fax: +49 761 49055-70 info@ptw.de www.ptw.de D850.131.05/01 en...
Contents General Information Contents Illustrations Intended Use Safety Information Operating Manual The Components of the MP3-PL System Perspex Tank Moving Mechanism Positioning Device SCANLIFT Carriage and Reservoir Control Unit and Control Pendant Assembly of the System Assembling SCANLIFT, Positioning Device, Chamber Cable Holding Device and Perspex Tank...
Page 5
MP3-PL Therapy Beam Analyzer T41050 Contents Service Manual Cleaning Preventive Maintenance 10 Storage 11 Disposal of the Product Literature Appendix: China Electronic Industry Standard Compliance Index D850.131.05/01 en...
Figure 3: MP3-PL System with accessories for measurements with a measuring and a reference detector Figure 4: MP3-PL Perspex tank with junction box and moving mechanism (top: front view; below: rear view) Figure 5: Connections at the junction box Figure 6:...
MULTIDOS, the TBA CONTROL UNIT, the TBA synchrotron). CONTROL PENDANT, the PC and the ionization chambers. For filling or emptying the MP3-PL Therapy Beam Analyzers the SCANLIFT carriage and reservoir is used. NOTE The basic system consists of the following compo-...
MP3-PL Therapy Beam Analyzer T41050 Safety Information Safety Information This is the safety alert symbol. It is used to Safety Information alert the user to potential hazards. Obey all DANGER safety messages that follow this symbol to avoid possible bodily injury or equipment Operation in areas where an explosion hazard damage.
Page 9
MP3-PL Therapy Beam Analyzer T41050 Safety Information WARNING WARNING Liquids must not enter the product. If liquids Improper handling. have entered the product, it must be thor- Patient Hazard! oughly inspected before being used again. The product is a medical electrical device and ...
Page 10
MP3-PL Therapy Beam Analyzer T41050 Safety Information WARNING WARNING Electricity is a source of risk, particularly when Magnetic and electrical fields are capable of in- the product is not in perfect operating condition terfering with the proper performance of the de- or when it is operated inappropriately.
Page 11
Use the product only in conjunction with the shifted (e.g., laterally by 15 mm). Continued irra- products approved for this purpose by PTW diation through the original position can lead to and/or the products listed in the section "Intend- radiation damage of the entrance window mate- ed Use".
Page 12
If the water phantom is lowered to the bottom When used with the SCANLIFT, rotating the position by means of the SCANLIFT lift mecha- MP3-PL by means of the positioning device is nism, only very little space (approx. 10 –12 mm) not allowed.
MP3-PL Therapy Beam Analyzer T41050 Safety Information Symbols on the Nameplate NOTE Set up the device so that the operator has a Symbol Description clear, unobstructed view of the indicators and controls and can operate the controls easily. Please refer to user manual! NOTE The product bears the CE-mark.
MP3-PL Therapy Beam Analyzer T41050 Operating Manual – The Components of the MP3-PL System Operating Manual 1 The Components of the MP3-PL System Figure 2: MP3-PL System with accessories for measurements with 24 PinPoint chambers D850.131.05/01 en...
MP3-PL Therapy Beam Analyzer T41050 Operating Manual – The Components of the MP3-PL System Figure 3: MP3-PL System with accessories for measurements with a measuring and a reference detector D850.131.05/01 en...
MP3-PL Therapy Beam Analyzer T41050 Operating Manual – The Components of the MP3-PL System Perspex Tank The moving mechanism as well as several ancillary devices (refer to section 2) are attached to the Per- spex tank. The tank (Figure 4) is filled with water. The water level should remain approx.
MP3-PL Therapy Beam Analyzer T41050 Operating Manual – The Components of the MP3-PL System Moving Mechanism CAUTION Improper handling. Risk of injury! Do not reach into the Perspex tank while the moving mechanism is working. Pay attention that hairs or cloths of the user and other persons involved do not touch the rotating spindle.
MP3-PL Therapy Beam Analyzer T41050 Operating Manual – The Components of the MP3-PL System Positioning Device The positioning device allows the Perspex tank to be aligned, using the laser system of the accelera- tor and the lines on the walls of the tank.
MP3-PL Therapy Beam Analyzer T41050 Operating Manual – The Components of the MP3-PL System SCANLIFT Carriage and Control Unit and Control Reservoir Pendant The MP3-PL system is operated via the control unit CAUTION and the control pendant. Improper handling. Equipment damage!
MP3-PL Therapy Beam Analyzer T41050 Operating Manual – Assembly of the System 2 Assembly of the System Place the Perspex tank (6) on the base plate of NOTE the chamber cable holding device. Pay attention You can use the supplied tools to assemble the that the water connection of the Perspex tank (7) system.
MP3-PL Therapy Beam Analyzer T41050 Operating Manual – Assembly of the System Assembling the Junction 1 Bracket of the chamber cable holding device Box and the Moving 2 Screw M5x10 (16x) Mechanism 3 Base plate of the chamber cable holding device...
MP3-PL Therapy Beam Analyzer T41050 Operating Manual – Assembly of the System Check that the two bevel gear wheels engage properly in the centering shoulder and adjust, if necessary, by turning the threaded control shaft of the C-axis. ...
MP3-PL Therapy Beam Analyzer T41050 Operating Manual – Installation of the System for Measurement 3 Installation of the System for Measurement Align the Perspex tank by using the laser system NOTE of the accelerator. You can use the supplied tools to install the sys- tem.
MP3-PL Therapy Beam Analyzer T41050 Operating Manual – Installation of the System for Measurement Mounting Detectors on the Place the holder T21003.1.020 on a flat base. Alternatively mount holder Moving Mechanism T21003.1.020 on the robot’s table top using the two M6x35 screws and the washers.
MP3-PL Therapy Beam Analyzer T41050 Operating Manual – Installation of the System for Measurement 3.2.2 Linear Detector Holder with 24 PinPoint Chambers Type 31015 Place the holder T21004.1.003 on a flat base. Alternatively mount holder T21004.1.003 on the robot’s table top using the two M6x30 screws and the washers.
MP3-PL Therapy Beam Analyzer T41050 Operating Manual – Installation of the System for Measurement Vertical mounting refer to Figure 15 Mount the holder T21004.1.004 on the C-slider with the two M4x16 screws (4, Figure 15). Mount the linear detector holder with the Pin- Point chambers on the holder T21004.1.004 us-...
MP3-PL Therapy Beam Analyzer T41050 Operating Manual – Installation of the System for Measurement Horizontal mounting refer to Figure 16 Mount the linear detector holder with the Pin- Point chambers on the holder T21004.1.004 us- ing the two M4x20 screws (4, Figure 16).
MP3-PL Therapy Beam Analyzer T41050 Operating Manual – Installation of the System for Measurement 3.2.3 Bragg Peak Chamber Type 34073 and Use following holders for the Bragg Peak cham- Bragg Peak Chamber Type 34070 bers: Bragg Peak chamber Type 34073 with holder CAUTION T41029.1.200 and...
MP3-PL Therapy Beam Analyzer T41050 Operating Manual – Installation of the System for Measurement Connecting the Detectors Bundle the chamber cables with the supplied Velcro strips as shown in Figure 19. 3.3.1 24 PinPoint Chambers Type 31015 with ...
MP3-PL Therapy Beam Analyzer T41050 Operating Manual – Installation of the System for Measurement Positioning of the Chamber Cable Holding De- vice for Measurements: CAUTION Improper handling. Risk of injury! The chamber cable holding device is very heavy (several kilograms).
MP3-PL Therapy Beam Analyzer T41050 Operating Manual – Installation of the System for Measurement Mounting and Connecting 3.3.2 Bragg Peak Chamber Type 34073 the Reference Detector Attach the plug holder T4316/U161 to the Per- spex tank. If required, connect the ionization chamber to the extension cable T2954/K2-6.
MP3-PL Therapy Beam Analyzer T41050 Operating Manual – Installation of the System for Measurement Insert the reference detector into the holder as Connect the reference detector to detector input shown in Figure 21 as far as it will go. Make labeled CHANNEL 2 on the rear side of the sure the side labeled "Focus"...
MP3-PL Therapy Beam Analyzer T41050 Operating Manual – Installation of the System for Measurement Mounting the Wedge Mounting the Film Holding T41029.1.220 Device T40040 Figure 25: Mounting the wedge T41029.1.220 Figure 26: Mounting film holding device T40040 1 Screw M6x16 (4x) 1 Screw M6x20 (4x) 2 Wedge T41029.1.220 with holder...
MP3-PL Therapy Beam Analyzer T41050 Operating Manual – Installation of the System for Measurement Connecting SCANLIFT 3.9.3 Disconnecting SCANLIFT from Perspex Tank 3.9.1 Preparing SCANLIFT CAUTION refer to Figure 27 Improper handling. Equipment damage! Put the hose clip onto the open tubing end of the water hose on the top of the SCANLIFT.
Perspex tank with water up to the top horizontal energy radiation. line, the pumping process has to be started again and stopped manually. PTW recommends the following procedure for initial filling of the system: Pumping from SCANLIFT Water Reservoir into ...
Page 43
MP3-PL Therapy Beam Analyzer T41050 Operating Manual – Installation of the System for Measurement Pumping from the Perspex Tank into SCANLIFT First empty the Perspex tank by draining the Water Reservoir water from the Perspex tank into the SCANLIFT water reservoir.
MP3-PL Therapy Beam Analyzer T41050 Operating Manual – Installation of the System for Measurement 3.11 Connecting the Control Unit Front Panel and Control Pendant CONTROL PENDANT socket: connection for TBA CONTROL PENDANT CAUTION Direct or scattered radiation NOTE Equipment Damage!
MP3-PL Therapy Beam Analyzer T41050 Operating Manual – Installation Check 4 Installation Check Before starting the MP3-PL system, please check When the devices, e.g. TBA CONTROL UNIT and the following points carefully: MULTIDOS, have been switched on, you will get the following indications: ...
MP3-PL Therapy Beam Analyzer T41050 Operating Manual – Error Messages and Troubleshooting 5 Error Messages and Troubleshooting If problems occur, please check the installation as NOTE outlined in section 2. The following list of possible When contacting our Service team, please give problems may be a useful guideline for trouble- the full serial number of your device.
MP3-PL Therapy Beam Analyzer T41050 Technical Manual – Technical Specifications Technical Manual 6 Technical Specifications Only values with specified tolerances or limits are guaranteed. Values without tolerances are for information purposes only. Description MP3-PL Therapy Beam Analyzer, T41050 Manufacturer PTW-Freiburg...
Page 48
MP3-PL Therapy Beam Analyzer T41050 Technical Manual – Technical Specifications Moving mechanism Moving range A-axis 500 mm Moving range B-axis 408.5 mm Moving range C-axis 500 mm Reproducibility 0.1 mm Positioning accuracy 0.5 mm Step interval 0.1 mm Max. cable length...
WARNING In addition, ensure that all other service tasks are Liquids have entered the product. only performed by PTW-Freiburg or by a person Shock Hazard, Erroneous Measurement! authorized by PTW-Freiburg. The device must be thoroughly inspected by service personnel before being used again.
Page 52
Service Manual – Cleaning Device Surfaces Perspex Tank First of all, disconnect the MP3-PL system from The Perspex tank is drained by pumping the water the power line, before attempting to clean the into the SCANLIFT water reservoir. device surfaces.
It is recommend- ed to perform a technical inspection of the measur- Check before each use ing system every 2 years at PTW-Freiburg or a qualified calibration laboratory. Before each use, visually inspect the device and the cables for signs of mechanical damage.
10 Storage 11 Disposal of the Product To preserve a constant evenness of the plates The typical lifetime of the MP3-PL Therapy Beam made of PMMA (e.g. holder for inhomogeneous Analyzer system is about 10 years. phantom), pay attention that these plates are stored...
MP3-PL Therapy Beam Analyzer T41050 Literature Literature Council Directive 93/42/EEC concerning medical devices (Medical Device Directive - MDD) Council Directive 97/43/EURATOM on health protection of individuals against the dangers of ionizing radiation in rela- tion to medical exposure IEC 61010-1 Safety requirements for electrical equipment for measure- ment, control and laboratory use;...
MP3-PL Therapy Beam Analyzer T41050 Appendix: China Electronic Industry Standard Compliance Appendix: China Electronic Industry Standard Compliance 附录:中华人民共和国电子产业标准 This Supplement concerns China Electronic Industry Standard Compliance 此附录涉及中华人民共和国电子产业标准的相关规定 The following product pollution control information is provided according to SJ/T11364-2006 Marking for Control of Pollution caused by Electronic Information Products.
Page 57
Data listed in the table represents best information available at the time of publication. This table shows where these substances may be found in the supply chain of PTW electronic information products, as of the date of sale of the enclosed product. Note that some of the component types listed above may or may not be a part of the enclosed product.