The product fully complies with the require- modification or repair is carried out by ments of the following standards: PTW-Freiburg or by persons authorized by IEC 60580 PTW-Freiburg, and if the product is used in compliance with the technical documentation. − Electrical safety −...
Page 3
DIAMENTOR E2 General Information − PTW-Freiburg works in strict accordance with a PTW-FREIBURG Physikalisch-Technische Werkstätten quality management system which is continu- ously updated according to national and inter- Dr. Pychlau GmbH national standards. Lörracher Str. 7 79115 FREIBURG GERMANY Phone: +49 761 49055-0...
Table of Contents INTENDED USE FOR YOUR SAFETY OPERATING CONTROLS AND INDICATORS FIRST TIME OPERATION Setting up the DIAMENTOR E2 Connecting the Supply Voltage Connecting the Chamber and Printer and Putting into Operation Mounting the ionization chamber Basic Device Settings (Setup Menu)
The DIAMENTOR E2 is a single channel micro- processor-controlled measuring system. WARNING The DIAMENTOR E2 is equipped with a serial con- nector (socket SERIAL) to connect a label printer or indicates a hazard. If not avoided, the hazard a PC.
Page 6
DIAMENTOR E2 For Your Safety Safety Information − Liquids must not enter neither the display DANGER unit nor the adapter. If liquids have entered Explosion Hazard – The product is not de- the product, it must be thoroughly inspected signed for use in areas of medical locations before being used again.
Page 7
− The DIAMENTOR E2 should not be used in manufacturer of the other equipment or an the immediate vicinity of or placed on top of expert about the effects of the connection on or below other equipment.
Page 8
Equipment damage - If you move the X-ray into the DC supply, the system is powered, unit do not bump into doorframes or walls to even if the DIAMENTOR E2 is turned off. avoid damage to the DIAMENTOR E2. D651.131.00/05 en...
Page 9
Definition of patient environment caused by Electronic Information Products (ACPEIP)“ (China RoHS) (refer also to Appendix) The connection of the DIAMENTOR E2 to the ground wire does not have any protection func- tion here, this serves as function earth. D651.131.00/05 en...
Page 10
DIAMENTOR E2 For Your Safety Information for Installation Inside the patient Outside the patient Not medical used area environment environment Power Power supply unit limiter T99052 T26060 Printer / PC IEC 60950-1 DIAMENTOR E2 T11033 Printer / PC IEC 60950-1...
DIAMENTOR E2 Operating Controls and Indicators 3 Operating Controls and Indicators Figure 3: DIAMENTOR E2 Socket SERIAL for connection of label printer or PC (RS232 interface; RJ12 Western Plug) Folding stand Connection ionization chamber (connecting system A ) Connection adapter...
DIAMENTOR E2 WARNING The system has no means of isolation from the The DIAMENTOR E2 can be set up by means of supply voltage. When the mains plug is con- the folding stand on the back of the device. nected to the power line or the cable is clamped...
Page 13
ANSI/AAMI ES 60601-1/ Installation in USA - In case of a 240 VAC- CAN/CSA-C22.2.No.60601-1. line voltage the DIAMENTOR E2 must only be The DIAMENTOR E2 may be operated only connected to center tapped circuits. with power limiter T26060. Extension cords must not be used.
Page 14
CAN/CSA-C22.2.No.60601-1 by the end user. Figure 5) • To achieve the requirements for electro- The DIAMENTOR E2 may be operated only with magnetic compatibility it is recommended to power limiter T26060 in compliance with the follow- connect the DC power supply output to GND ing conditions.
• Always turn off the device before connecting or General information removing peripheral devices. The DIAMENTOR E2 can be operated with every PTW-DIAMENTOR chamber featuring connecting • Connect the chamber to 3 (refer to Figure 3). system A (Figure 3, item 3).
DIAMENTOR E2 First Time Operation 4.5 Basic Device Settings (Setup Entering the chamber calibration factor Menu) Check the setting for the chamber calibration factor. The original chamber factor K is indicated on the General information chamber housing. The setup menu will be activated by pressing the...
Display of the dose area rate in µGym²/s. This mode is indicated by a square at the left margin of the display. The DIAMENTOR E2 will conduct an internal con- stancy check as soon as the TEST button is „Dose“ measuring mode (Setup: doSE) pressed for 2 seconds.
The print mode can be switched on / off (display indicated / determined value of the Prn on / Prn off) over setup menu (refer to sec- DIAMENTOR E2 must be multiplied with a correc- tion 4.5). tion factor to the air density correction.
HU-Err Chamber voltage missing, device is un- Disconnect the chamber cable from the able to measure DIAMENTOR E2. When the message continues appearing the device is defec- tive and must be repaired. When the message clears the chamber cable is defective and must be replaced.
− Enter the new chamber calibration factors as Electrical Calibration described in section 4.5. For this, multiply the The DIAMENTOR E2 is electrically calibrated in the chamber calibration factor K by a correction factory. You can verify the calibration as described factor of 1.10 for over-couch tubes and of...
Page 21
− Place the measurement stand without detector − Note down the dose area product reading on the patient couch (phantom at the top). shown on the DIAMENTOR E2 (DAP Measured − Start the fluoroscopy and determine the field − Calculate...
Before connecting a PC via the RS232 inter- Switch on the printer first or simultaneously with the face, set the print mode to “Prn off” (see sec- DIAMENTOR E2. Set the print mode to “Prn on” tion 4.5). (see section 4.5).
First of all, disconnect the devices from the Repair power line, before attempting to clean or disin- fect them. Repairs may only be carried out by PTW-Freiburg or by persons / companies authorized by • Rub the devices down with a moist cloth; no PTW-Freiburg.
9.3 Disposal of the Product Safety inspections Only regularly inspected product s are fail-safe. To The typical lifetime of the DIAMENTOR E2 is about preserve the functional security and operational 10 years. safety, the safety inspections are to be executed at regular intervals according to national regulations.
10 Technical Specifications Only values for which tolerances or limits are specified are guaranteed. Values for which no tolerances are specified are for information only. DIAMENTOR E2 T11033 Device designation and use dose area product meter according to IEC 60580 for measurements...
Page 26
DIAMENTOR E2 Technical Specifications + 300 VDC ± 10 V Chamber voltage < 0.1 mA Current limitation of the chamber volt- 24 VDC ± 10 % Power 3 W max. Power input II according to IEC 60601 Protection class Suitable for use in an electromagnetic environment according to...
Call up the Service menu by pressing the TEST and T25030-3 RS232 interface cable PRINT buttons simultaneously. In the Service menu L104078 Wall mount for DIAMENTOR E2 mode the buttons are used for cursor control, to on request DIAMENTOR chamber change and confirm settings and to quit a menu.
Page 28
Harmonic emissions complies poses. IEC 61000-3-2 Voltage fluctuations / complies flicker emissions IEC 61000-3-3 The DIAMENTOR E2 fulfils the requirements of IEC/CISPR 11 Class B, anyhow according to its intended use it is not for operation in domestic establishments. D651.131.00/05 en...
Electromagnetic Compatibility (EMC) according to IEC 60601-1-2 Guidance and Manufacturer’s Declaration – Electromagnetic Immunity The DIAMENTOR E2 is intended for use in the electromagnetic environment specified below. The customer or the user of the DIAMENTOR E2 should assure that it is used in such an environment.
Page 30
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity The DIAMENTOR E2 is intended for use in the electromagnetic environment specified below. The customer or the user of the DIAMENTOR E2 should assure that it is used in such an environment. Immunity Test IEC 60601 Compliance Electromagnetic Environment –...
Page 31
Recommended separation distances between portable and mobile RF communications equipment and the DIAMENTOR E2 The DIAMENTOR E2 is intended for use in an electromagnetic environment in which radiated RF distur- bances are controlled. The customer or the user of the DIAMENTOR E2 can help prevent electromagnetic...
DIAMENTOR E2 Appendix: China Electronic Industry Standard Compliance / Special Adjustments Appendix: China Electronic Industry Standard Compliance 附录:中华人民共和国电子产业标准 This Supplement concerns China Electronic Industry Standard Compliance 此附录涉及中华人民共和国电子产业标准的相关规定 The following product pollution control information is provided according to SJ/T11364-2006 Marking for Control of Pollution caused by Electronic Information Products.
Page 33
• Data listed in the table represents best information available at the time of publication. • This table shows where these substances may be found in the supply chain of PTW electronic information products, as of the date of sale of the enclosed product. Note that some of the component types listed above may or may not be a part of the enclosed product.