Service Manual webline UNIDOS 2006-06 Se/Sa D776.133.0/1...
Page 2
webline UNIDOS Contents Contents For Your Safety Spare Parts List Regularly Maintenance Checkups (Function tests) Constancy Test Maintenance Further Tests System Information Troubleshooting Functional units Front panel unit Measuring Module Main Board Service Software D776.133.0/1...
UNIDOS user manual and is valid only in Description PTW ID No. or conjunction with that manual. Order Designation The safety statements appearing in the indi- Desiccator cartridge L137004 cated user manual are applicable and must be...
Page 4
The UNIDOS must be inspected at regular intervals according to national regulations at PTW- Freiburg or a qualified calibration laboratory. Safety inspections and inspections of measurement technique must also be performed after repairs or after each influence which may have changed the behavior of the device.
webline UNIDOS Regularly Maintenance 3.2 Constancy Test The following tests should be executed regularly. They make it possible to recognize slow alterations of the characteristics of device or detector. Test Interval Accessories Comment Detector test weekly HV check at detector; only range Low Device test weekly checks device only;...
UNIDOS Regularly Maintenance UNIDOS webline SERVICE 2006-02-01 12:04:09 3.4 Further Tests System information Ethernet-interface Device name: UNIDOS webline The Ethernet interface can be checked using the Type number: T10021 VNC Viewer. Serial number: 5 Hardware revision: A02 Firmware: 1.12-1...
Opening of the device and working at the device the procedures to apply, if problems occur during must be performed only by PTW-Freiburg or by the operation of the device or error messages are persons authorized by PTW-Freiburg. shown on the display. The particular sections are: WARNUNG 10.1...
webline UNIDOS Functional units In case of exchanging the front panel unit all these data have to be saved first and later on entered into the new front panel unit by means of the service software. The main CPU contains also the real-time clock and the associated back-up battery.
webline UNIDOS Functional units 5.3 Main Board Supply voltages Wide Power Mains range supply Connector & units AC / DC Filter Rear panel Charge control Voltages measurement Rechargeable batteries Electrical Isolation Host I/O (SERIAL) Ethernet TRIP Output Relay Figure 3: Main board The main board contains the individual subunits for generation and monitoring of the supply voltages as well as for charging the rechargeable batteries.
PC − Update of the device firmware − Restore of the device settings as well as the detector data base If a firmware update from PTW-Freiburg is released webline for the UNIDOS , the user can update the firm- ware version on the newest version by means of this software.
Need help?
Do you have a question about the UNIDOS webline and is the answer not in the manual?
Questions and answers