Saniclave Front-Loading Series
INSTALLATION INSTRUCTIONS
Remove Saniclave from packaging and place it on a dry level surface. Plug it into
an accessible 120 VAC power receptacle. For optional printer setup, see pg. 6.
The Saniclave is now ready for use. See operating instructions on pg. 7.
Upon receiving the Saniclave, it is important that you follow these simple steps:
1. Examine carton and contents for possible damage to the product caused
during shipping. If damage is present contact the carrier at once. Failure to do
so immediately may release the carrier from any liability of damage.
2. Fill out the checklist (see pg. 14). If the shipment is incomplete, please
contact us at support@revsci.com or (800) 775-7942 x 2.
3. Carefully read this instruction manual before operating the Saniclave.
4. 4. Visit the product registration webpage provided on pg. 12 by scanning
the QR code or entering the URL manually and complete the registration
online.
INTENDED USE
The Revolutionary Science Saniclave is designed to be used in medical and dental
clinics, hospitals and other facilities where reusable sterile equipment is used. It
is intended to sterilize wrapped and unwrapped heat and moisture stable solid
instruments, mated surfaces, knurled and hinged devices (excluding lumened devices
and dental hand pieces) that are compatible with saturated steam sterilization at
121 degrees C° for 30 minutes.
SAFEGUARDS & CAUTIONS
WARNING! Do not use this autoclave in an explosive or oxygen rich atmosphere.
Failure to do so may result in serious injury or death.
CAUTION: Chamber contains hot water and steam during and after operation.
Hot water may drip from cover during removal and can cause scalding.
Occasionally parts such as the door seal, control panel decal, and ZPD
should be replaced after heavy usage: These easily replaceable parts are
available through Revolutionary Science. No tools are required.
17319 Lake Blvd. Shafer, MN 55074 | 800-775-7942 | Fax: 775-218-7015 | support@revsci.com | revsci.com