MR Safe
MR Conditional
14
Trigger converter
box
Active driver
Passive driver

Labeling

2.3.1
ASTM International developed a new classification system for
implants and ancillary clinical devices. The following definitions
apply:
An item that poses no known hazards in all MR environments. MR Safe items
include nonconducting, nonmagnetic items such as a plastic petri dish. An
item may be determined to be MR Safe by providing a scientifically based
rationale rather than test data.
An item that has been demonstrated to pose no known hazards in a specified
MR environment with specified conditions of use. Field conditions that define
the specified MR environment include field strength, spatial gradient, dB/dt
(time rate of change of the magnetic field), RF fields, and SAR. Additional
conditions, including specific configurations of the item, may be required.
MR Conditional devices (for example, RF communications equipment) may
present hazards as well. Please observe the manufacturer's operator manual
to avoid potential hazards and injuries.
MR Elastography | Operator Manual
Print No. MR-06020G.640.08.04.02