IV - Maintenance
1

PREVENTIVE MAINTENANCE

Contact your QUANTEL MEDICAL or your local distributor at least every six months to carry out a preventive
maintenance visit to confirm the correct function of your system.
At each six-monthly visit QUANTEL MEDICAL or your local distributor will clean the optics, check the safety
filter coatings, and check general performance, alignment and calibration of the system.
At least once a year QUANTEL MEDICAL or your local distributor will fully re-calibrate the system and check
ground resistance and earth leakage current (in accordance with IEC 60601-1).
2

CARE AND MAINTENANCE OF THE OPTIMIS FUSION LASER SYSTEM

2.1

CLEANING AND DISINFECTION FOR PATIENT AREAS

The use of the OPTIMIS FUSION involves only brief skin contact with the patient, but attention should be
given to the possibility of cross-contamination between patients via the patient contact areas of the system.
These are the chin rest, headrest and steadying handles.

2.1.1 RESPONSIBILITIES FOR CLEANING AND DISINFECTION

The health care facility where the instrument is to be used is responsible for:
1)
Determining the level of cleaning and disinfection of patient contact areas required between patients.
2)
Appropriate education and training for staff required to carry out cleaning and disinfection.
3)
Ensuring that routine cleaning and disinfection methods used in the facility are compatible with the
instrument.
4)
Routine cleaning of the entire system.
2.1.2 CLEANING AND DISIFECTION OF PATIENT AREAS
Although cleaning and disinfection procedures and standards within different health care facilities may vary
considerably, the following points are provided as general guidelines:
1)
Thorough cleaning of all patient contact areas is recommended for each procedure.
2)
Disposable chinrest papers can also be attached to the chinrest area and changed between patients.
3)
Manual cleaning can be carried out by wiping all contact areas using a suitable liquid cleaning agent
which is non-corrosive, non-toxic and Iow in residue.

2.1.3 CLEANING THE GONIOSCOPY CONTACT LENSES

The contact lenses for use with the OPTIMIS FUSION have a special low-reflectivity coating bonded to
the lens, and they must therefore be handled carefully.
As soon as a lens is removed from the patient's eye, thoroughly rinse it in cold or warm water to remove
salts, mucous and gonioprism solution. Wash in warm water with a few drops of clear dishwashing liquid,
then rinse with cool water and blot dry. Dry completely before storing in the case.
To effectively disinfect the lens, soak in a solution of sodium hypochlorite 1,2° chl. for a recommended
exposure time of 30 minutes. Rinse thoroughly with cool water and blot dry. Dry completely before storing
in the case.
To sterilize, use ethylene oxide gas with aeration not to exceed 125°F (52°C), following the sterilizer
manufacturer's recommendation. Remove the lens from the case before sterilizing.
CAUTION:
Do not boil or autoclave a lens. Do not use alcohol, hydrogen peroxide or acetone on a lens, as this will
damage it and render it unusable.
XL_FUSION_ME_AN_160314
1

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the OPTIMIS FUSION and is the answer not in the manual?

Questions and answers

Dushmantha
April 5, 2025

What is E 4 error

Summary of Contents for Quantel Medical OPTIMIS FUSION

  • Page 1 CLEANING AND DISINFECTION FOR PATIENT AREAS The use of the OPTIMIS FUSION involves only brief skin contact with the patient, but attention should be given to the possibility of cross-contamination between patients via the patient contact areas of the system.
  • Page 2 Use one tissue or one Q-tip per wipe, then discard and use a fresh one for the next wipe. CAUTION In case any scratching is visible do not use the laser system and call QUANTEL MEDICAL or your local distributor.
  • Page 3 Attach a piece of photographic paper on the alignment target / focusing bar (See Figure 1) Figure 1: Piece of photographic paper in position Switch ON the main switch located at the back of the OPTIMIS FUSION unit. The slit lamp is automatically turned “ON” and illuminates the target paper.
  • Page 4 User manual IV – Maintenance WARNING If the aiming beam is not centered: do not use the système and contact QUANTEL MEDICAL Service Department or your local distributor. Adjust: - Burst mode to 1 pulse - Focus offset to 0 (30µm offset)
  • Page 5 Attach a piece of photographic paper on the alignment target / focusing bar (See Figure 1 in the above section 2.2.3) Switch ON the main switch located at the back of the OPTIMIS FUSION unit. The slit lamp is automatically turned “ON” and illuminates the target paper.