Hanna Instruments HI981204 - Waterproof ORP And Temperature Tester Manual

Hanna Instruments HI981204 - Waterproof ORP And Temperature Tester Manual

Advertisement

Hanna Instruments HI981204 - Waterproof ORP And Temperature Tester Manual

Introduction

Thank you for choosing a Hanna Instruments product. Please read this instruction manual carefully before using the tester. For more information about Hanna Instruments and our products, visit www.hannainst.com or e-mail us at sales@hannainst.com.

For technical support, contact your local Hanna Instruments Office or e-mail us at tech@hannainst.com.

Preliminary Examination

Remove the tester and accessories from the packing material and examine it carefully. If you require any further information, please contact Hanna Instruments technical support team at tech@hannainst.com.
HI981204 is delivered in a cardboard box and is supplied with:

  • HI73120 ORP electrode
  • HI700224 ORP test solution, 470 mV, 20 mL sachet (6 pcs.)
  • HI73128 Electrode removal tool
  • Storage / Protection cap
  • 1.5V batteries (4 pcs.)
  • Instrument quality certificate
  • Instruction manual

Note: Save all packing material until you are sure that the tester works correctly. Any damaged or defective item must be returned in its original packing material with the supplied accessories.

General Description & Intended Use

HI981204 is a compact ORP and temperature tester, part of Hanna Instruments pool-line family. It features a twobuttons operation system and is easy to use. The compact and waterproof casing is designed to float if accidentally dropped in the water. The tester is factory calibrated and measures temperature in both °C and °F. It displays a stability tag that will disappear once the reading has stabilized.

Battery level and low-battery indicator are displayed on the LCD to alert the user in the event that low battery power could adversely affect the readings.

BEPS (Battery Error Prevention System)
The tester will automatically shut off if there is not enough power to guarantee an accurate reading.

ORP Probe
HI981204 is supplied together with HI73120 replaceable ORP electrode.

Temperature Sensor
The stainless steel temperature sensor facilitates faster and more accurate temperature measurement.

Specifications

Range ±1000 mV
-5.0 to 60°C (23.0 to 140.0°F)
Resolution 1 mV
0.1°C (0.1°F)
Accuracy
@ 25°C (77°F)
±2 mV
±0.5°C (±1.0°F)
Calibration Factory calibrated
Electrode HI73120 ORP electrode
Battery type 1.5 V
Battery life Approximately 300 hours of use
Auto-off After 8 minutes
Environment -5 to 50°C (23 to 122°F); RH 100%
Dimensions 171 x 41 x 26 mm (6.7 x 1.6 x 1.0")
Weight 83 g (2.9 oz.)

Functional Description

Functional Description

LCD Display

LCD Display

  1. Stability tag
  2. Battery life percentage indicator
  3. Low battery indicator
  4. Measurement unit
  5. First LCD line
  6. Second LCD line

Operational Guide

Turning the Tester ON / OFF

Press and hold /MODE. All LCD segments will be displayed for a few seconds followed by battery life percentage indicator (e.g. % 100 BATT).

To turn the tester OFF, from measurement mode, press / MODE.
"OFF" will be displayed on the second LCD line and the tester will turn off.

HOLD Mode

From measurement mode, press and hold SET/HOLD until HOLD is displayed. mV reading will be frozen on the LCD.
Press either button to resume measurement mode.

Setup

Setup mode allows the selection of temperature unit.
To enter the Setup mode, press ⏻/MODE until "TEMP" and the current temperature unit are displayed on the second LCD line (e.g. "TEMP °C").
Use SET/HOLD to change the temperature unit, and press ⏻/ MODE to return to measurement mode.

Measurement

Place the electrode in the solution to be tested while stirring it gently. The measurement should be taken after the stability tag ( ) has disappeared.

The ORP (mV) value is displayed on the first LCD line and the temperature on the second LCD line.

Note: If measurements are taken successively, rinse the probe thoroughly in distilled or deionized water to eliminate cross-contamination. After cleaning, rinse the probe with some of the sample to be measured.

Maintenance

Please read the information below to ensure the highest possible accuracy.

  • If measurements are taken successively, rinse the probe thoroughly in distilled or deionized water to eliminate cross-contamination.
  • Clean the electrode monthly using ORP cleaning solution.
    Keep the electrode in HI70614 General purpose cleaning solution for 30 minutes and rinse the probe thoroughly with water.
  • When not in use, add a few drops of HI703004 Storage solution to the protective cap. Never store the probe in distilled or deionized water.

Electrode Replacement

Use supplied removal tool (HI73128) to replace the ORP electrode.

  1. Insert the tool into the probe cavity.
    Electrode Replacement - Step 1 - Insert the tool
  2. Rotate the tool counterclockwise.
    Electrode Replacement - Step 2 - Rotate the tool
  3. Pull the electrode out by using the other side of the tool.
    Electrode Replacement - Step 3
  4. Insert a new ORP electrode following the instruction steps in reverse order.

Battery Replacement

Battery life percentage indicator is displayed at power on. If the battery level drops below 5%, the symbol lights up indicating that the batteries should be replaced soon.

If the battery level is not adequate to guarantee an accurate reading, the tester automatically turns off.

To replace the batteries, follow the next steps:

  1. Turn OFF the tester.
  2. Remove the four screws on the top of the tester to open the battery compartment.
  3. Remove the old batteries.
  4. Insert four new 1.5V batteries in the battery compartment while paying attention to the correct polarity.
  5. Close the battery compartment using the four screws.

Note: Only use the battery type specified in the manual. Old batteries should be disposed in accordance with local regulations.

Accessories

Electrodes
HI73120 Replaceable ORP electrode
HI73128 Electrode removal tool
ORP Test Solution
HI7021M 240 mV @25°C, ORP test solution, 230 mL
HI70224L 470 mV @25°C, ORP test solution, 500 mL
Electrode Cleaning Solution
HI70614L General purpose cleaning solution, 500 mL
Electrode Storage Solution
HI703004L Electrode storage solution, 500 mL
Pretreatment Solution
HI7091L Pretreatment reducing solution
HI7092M Pretreatment oxidizing solution, 230 mL
Other Accessories
HI740026P Replacement 1.5 V batteries (12 pcs.)

Documents / Resources

References

Download manual

Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.

Download Hanna Instruments HI981204 - Waterproof ORP And Temperature Tester Manual

Advertisement

Need help?

Need help?

Do you have a question about the Pool Line HI981204 and is the answer not in the manual?

Questions and answers

Table of Contents