The lithium-ion rechargeable battery
in the UroShield
disassembled, heated above 100 degrees
Celsius, incinerated, or exposed to water.
Be aware of the actuator cable which can
get caught, entangled or even wrapped
around a person's body and lead to
potential injury or restrict blood flow.
Use of UroShield
with other equipment should be avoided
as it could result in improper operation.
If such use is necessary, observe the
UroShield
®
operating normally.
Use of accessories, transducers, and cables
other than those specified or provided
by NanoVibronix could result in increased
electromagnetic emissions or decreased
electromagnetic immunity
of this medical device and result in
improper operation.
P ortable RF communications equipment
(including peripherals such as antenna
cables and external antennas) should not
be no closer than 30 cm (12 inches) to
any part of the UroShield
operation. The performance of
UroShield
®
R efer to Appendix A for information
on maintaining basic safety and
essential performance with regard to
electromagnetic disturbances
of UroShield
The UroShield
therefore, should be removed from the
catheter before entering the MRI suite.
driver must not be
®
adjacent to or stacked
®
system to verify it is
may be affected.
.
®
is not MRI compatible and
®
system during
®