THE DRIVER

The life of the driver is dependent on the
battery use, refer to section 5.2. To clean
the driver, use 70/30 Iso Propyl Alcohol or
disinfectant medical wipes.
Do not use solvents such as acetone as
they may damage the product. Do not
immerse the UroShield
any other liquid.
T he driver is flame resistant according to
UL-94HB. It does not contain flammable
materials and will not accelerate a fire.
The driver is not intended for use in the
presence of flammable liquids.
A faulty unit which is still under warranty,
can be sent to NanoVibronix for
replacement. Refer to Appendix D for
conditions of warranty.
The driver does not contain recyclable
material.

THE ACTUATOR

The actuator is intended for single use
with one catheter only and should be
changed together with the catheter
(Maximum 30 days).
If the actuator becomes soiled, it can be
cleaned with a damp sponge and then
allowed to dry.
After use, the actuator must be disposed
of in accordance with the relevant medical
facility's standards for the disposal of
used medical equipment.
driver in water or
®
13
6.1
6.2
6