Prime Controls LH200 Operating Instructions Manual page 5

Light leak detector

Advertisement

The result flags are D, N and L. Their meaning is as follows:
D – the average dark current exceeded the dark limit
N – the dark spread exceeded the spread limit
L – the peak value exceeded the leak limit
S
– Spread Limit
Displays and opens the spread limit value for change. If the displayed value is to be retained, simply press
ENTER. Otherwise, enter the new value then press ENTER. Acceptable range is from 1 to 4000. The
default is 500. This parameter determines how much variation in signal is allowed during the dark interval of
the test cycle. If the spread in dark readings during a test cycle exceeds the dark spread, the report LED is
set to amber and the RESULT output is turned off (reject state). For most applications this test has limited
meaning and may be essentially disable by setting the limit to a high value. Its primary usage is for detecting
electrical noise.
V
– Version
Displays the controller firmware version number in the form of 1.01 for example.
?
– Menu
Displays the command menu as shown at the beginning of this section.