Prime Controls SD220 Operating Instructions Manual

Shell/tab detector
Hide thumbs Also See for SD220:

Advertisement

OPERATING INSTRUCTIONS
MODEL SD220 SHELL/TAB DETECTOR
DESCRIPTION
The Model SD220 Shell Detector for conversion press applications is a rugged but
sensitive dual channel instrument designed specifically to detect and report missing
blanks or double blanks at the infeed to a conversion press and to detect and report
ends that exit the press without tabs.
A complete system for one lane of a press comprises a control module housed in an
aluminum enclosure and four probes. One transmitting and receiving probe pair
senses doubles at the press input while the second transmitting and receiving pair
senses the presence or absence of tabs at the press output.
March 13, 2019

Advertisement

Table of Contents
loading

Summary of Contents for Prime Controls SD220

  • Page 1 MODEL SD220 SHELL/TAB DETECTOR DESCRIPTION The Model SD220 Shell Detector for conversion press applications is a rugged but sensitive dual channel instrument designed specifically to detect and report missing blanks or double blanks at the infeed to a conversion press and to detect and report ends that exit the press without tabs.
  • Page 2: Table Of Contents

    DOCUMENT APPLICABILITY…………………………………………….……. Page 12 ELECTRICAL SPECIFICATIONS……………………………………………… Page 13 CONTROL MODULE DIMENSIONS…………………………………………... Page 14 WIRING RECOMMENDATIONS………………………………………………..Page 15 OUTPUT SIGNAL DIAGRAMS………………………………………………….Page 16 SUMMARY OF DIFFERNCES BETWEEN SD200 AND SD220…………… Page 17 SD200 TO SD220 MIGRATION…………………………………………………Page 18 LIMITATION AND EXCLUSION OF WARRANTIES………………………….Page 19...
  • Page 3: Control Module

    Control Module The control module allows for fast and easy setup and for quick diagnosis of system errors or problems. Setup is achieved through the simple press of a push-button switch or an external contact closure. Faults are reported on digital displays on the control module and through FAULT outputs that may be connected to a PLC or system controller.
  • Page 4: Probes

    Probes The SD220 may be used with any P15 or P70 series two wire probes or with older model three wire probes including AV, AY, AZ, AZA, and CB probes. When used with older 3 wire probes, the white wire in the probe cable is not used and must remain disconnected.
  • Page 5: Fault Message Interpretation

    If both channels calibrate successfully, calibration is complete. The calibration process may also be initiated through an external switch or signal controlling the CAL+ and CAL- inputs to the SD220. Activation of this input performs the same function as pressing the CALIBRATE push-button switch for the corresponding channel.
  • Page 6: Mounting The Control Module

    Note: Prime Controls assures compatibility with PN: CBL101-XX cables and the cabling pre-attached to certain probes. Other types of cabling, and also, cable lengths above 30 meters (approx 100 ft) are not advised.
  • Page 7: Setting Switch Options

    AUX inputs may be wired to provide external control of the sensing mode of the SD220. Connect a sinking driver or contact to the AUX- terminal and connect the AUX+ terminal to the 24 volt supply. Connect a sourcing driver to the AUX+ terminal and connect the AUX- terminal to COM.
  • Page 8: Sensing Aluminum Ends

    Channel 2 Switch Options Switch ====================================================== Sense aluminum ends Sense steel ends Sense aluminum tabs Sense steel tabs Select fail-safe mode Select compatibility mode Fixed thresholds Adjustable thresholds Sourcing outputs Sinking outputs Enable tab profiling Disable tab profiling Not used Not used NOTE: The DIP switches are read only upon power-up of the unit.
  • Page 9: Enable Or Disable Tab Profiling

    Set Outputs for Failsafe or Compatibility When SW3 is on, the sourcing outputs of the SD220 provide the same logic levels as the outputs of older double sheet units such as the DS33 and DS35, allowing for quick and easy retrofit installations. When SW3 is off, the output states are defined to provide maximum protection against loss of connection between the shell detector and the controlling PLC.
  • Page 10: Set Display Direction

    1. The 0 selects normal display mode, 1 selects inverted mode.. 4) When the desired value is on the display, wait 5 seconds and the display reverts to displaying the signal value in the selected mode. The range of values available for the SD220 adjustable parameters are as follows:...
  • Page 11: Parameter Value Ranges

    0% to 90% for off-press sensing only (SW2 On) Special Functions The SD220 offers two sets of optically isolated inputs on each channel that provide added control over the unit. These are the remote calibration input terminals labeled CAL+ and CAL- and the sensing mode inputs labeled AUX+, AUX-.
  • Page 12: Determining Firmware Version

    The revision number is displayed directly on the digital display. The two channels of the SD220 are independent and may be running different versions of firmware. Each must be checked separately.
  • Page 13: Electrical Specifications

    Sinking Outputs: On voltage: 0 volts Off voltage: Load pullup dependent Max current: 50 mA Output Overload Protection: Self-resetting thermal fuse Input and Output Transient Protection: 30 volt transient absorber. Probe Cabling: Prime Controls PN: CBL101-XX, Belden 8761, or exact equivalent.
  • Page 14: Control Module Dimensions

    SD220 CONTROL MODULE DIMENSIONS (8.25) 7.94 (7.63) (.313) (6.25) 101.6 (4.00) 4.76 28.6 (.188) (1.125) 20.6...
  • Page 15: Wiring Recommendations

    Total cable lengths greater than 30 M (approx. 100 ft.) are not advised. Use of excessively long cables, or cables not specified by Prime Controls, may cause issues within the receiver and/or probe detection circuits of the controller.
  • Page 16: Output Signal Diagrams

    SD220 OUTPUT SIGNALS CHANNEL 1 - INFEED CHANNEL 2 - DISCHARGE No Tab/Tab Outputs - Failsafe Mode NO TAB NO TAB no end no end no end no end NO TAB NO TAB no end no end no end no end...
  • Page 17: Summary Of Differnces Between Sd200 And Sd220

    3. If your installation uses the AUX input to quickly change between calibrations on different shell materials, it is necessary with the SD220 to activate the AUX input on both channels independently. The AUX inputs are not slaved together as in...
  • Page 18: Sd200 To Sd220 Migration

    Configuration of the SD220 has been simplified as compared to the SD200. It is no longer necessary to remove the cover to change options. These changes include: 1. The DIP switches, now eight per channel instead of four per unit, are accessible through a window on the front panel.
  • Page 19: Limitation And Exclusion Of Warranties

    LIMITATION AND EXCLUSION OF WARRANTIES All goods purchased form PRIME CONTROLS, INC. shall be free from defects in materials, design and workmanship under normal conditions of use for one year from the date of shipment. THIS WARRANTY IS THE SOLE WARRANTY AND IS...