Page 1
Mya Joy Double Electric Breast Pump Instructions for use A meda.com || 866.992.6332 || ParentCare@ameda.com...
Page 2
The Ameda HygieniKit is a closed system that contains a diaphragm that keeps the air from the pump and tubing separate from the pump parts that are exposed to your breast milk, so your tubing stays dry. This diaphragm helps prevent milk backup into tubing.
• Prior to use, always inspect power adapter for damage or exposed wires. If damage and/or exposed wires are found, do not use power adapter and call Ameda Customer Service ParentCare at 1.866.992.6332. • NEVER operate electrical product: if it has a damaged cord or plug; if it is not working properly; if it has been dropped, damaged, or becomes wet.
Page 5
• If batteries are in pump base, remove batteries before using the power adapter. • DO NOT wrap cord around the body of the adapter. • Use ONLY the AC power adapter that is recommended for the Ameda Mya Joy Breast Pump. • DO NOT shorten tubing or modify the breast pump kit.
SAVE THESE INSTRUCTIONS 2. INTENDED USE The Ameda Mya Joy Breast Pump is intended to express and collect the mother’s milk from the breasts of a nursing woman for the purpose of feeding the collected milk to a baby. 3. CONTRADICTIONS FOR USE There are no known contradictions for use with this product.
CLEAN TUBING: It is necessary to wash the tubing only if condensation is present. If condensation is present in the tube, wash and dry it immediately or replace the tubing. If milk is present in the tubing, do not attempt to wash or clean the tubing. Instead contact Ameda Customer Service at 1.866.992.6332. HOW TO WASH TUBING: 1.
5.2 Assembling the HygieniKit ® Milk Collection System Once your pump parts are completely dry before use, wash your hands thoroughly and assemble the HygieniKit Milk Collection System (Pump Kit). ® Adapter Cap Diaphragm Breast Flange Tubing Valve Tubing Adapter Bottle HOLD by the rim or thicker PUSH the valve onto the...
Page 10
ATTACH and TIGHTEN the bottle or INSERT the diaphragm into the top attach milk storage bag. of the flange. It is important that the inside of the diaphragm is completely DRY to avoid moisture from forming in the tubing. ATTACH the adapter cap(s) by The tubing should still be CONNECTED pressing it onto the top of the to the adapter cap(s) and tubing adapter.
6. INSPECTING BEFORE PUMPING IMPORTANT: Before using the Ameda Mya Joy and HygieniKit system, check the parts for cracks, chips, tears, or wear. They can affect the function of the kit. If parts need replacing due to defects, please call your local distributor or the location where you purchased the pump. Do not use damaged parts.
NOTE: After two minutes, the pump will automatically switch to Expression Mode unless you manually adjust prior. C. SUCTION STRENGTH LEVEL Level of suction strength during pumping. The Ameda Mya Joy has 2-phase adjustable suction modes: – Stimulation Mode: 1 - 6 levels –...
7.4 To Begin Pumping STEP 1: CONNECT THE POWER ADAPTER Insert the power adapter into LEFT side of pump base. STEP 2: INSERT TUBING Insert the tubing into the adapter into RIGHT side of pump base, then turn slightly to the right while pushing in firmly. A.
Page 14
STEP 5: ADJUST VACUUM LEVEL Adjust the Stimulation Vacuum Level between 1 and 6 to your preferred comfort level. Never set to uncomfortable level. TIP: Start off with the suction low, and then gradually increase the suction setting to your comfort level. You may find that the level of suction that you are comfortable with varies from day to day, during the day and even during the same expression session.
7.5 How to Fit Your Flange Check breast flange fit, if pumping causes discomfort a larger or smaller breast flange size may be needed. GOOD FIT FLANGE TOO SMALL FLANGE TOO LARGE During pumping, your nipple During pumping, some or all During pumping, more areola moves freely in the breast flange of your nipple rubs against the...
8. CLEANING THE HYGIENIKIT ® BEFORE EVERY USE Wash hands with soap and water. Inspect and assemble clean pump kit. If your tubing is moldy, discard and replace immediately. Clean pump dials, power switch, and countertop with disinfectant wipes, especially if using a shared pump. AFTER EVERY USE Store milk safely.
9. MAINTENANCE The breast pump base does not require service or maintenance except for changing batteries, if applies. 9.1 Breast Pump Valve and Diaphragm Check Inspect the valves and diaphragms for tears or holes. Replace any damaged valves and diaphragms prior to pumping.
Page 18
(1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. Caution: changes or modifications to this device are not expressly approved by Ameda may void the user's authority to use the device.
90 days on other pump components except for breast pump valves. Ameda’s sole obligation under this limited warranty shall be to repair or replace, at Ameda’s option, any Product that is determined to be defective by Ameda and determined to be covered by this express limited warranty. Repair or replacement under this limited warranty is the sole and exclusive remedy of the First User.
12. TECHNICAL DATA 12.1 Specifications Product Description Electric Breast Pump Model Ameda Mya Joy Suction Stimulation mode: Dual-Kit: 40 to 100 mmHg Stimulation mode: Single-Kit: 80 to 160 mmHg Expression mode: Dual-Kit: 50 to 250 mmHg Expression mode: Single-Kit: 80 to 250 mmHg...
12.2 Electromagnetic Compatibility TABLE 1 Guidance and declaration of manufacturer-electromagnetic emissions The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that it is used in such an environment. Electromagnetic environment - Emissions test Compliance...
Page 22
TABLE 3 Guidance and declaration of manufacturer-electromagnetic immunity. (For home healthcare environment) The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that it is used in such an environment. Electromagnetic environment IMMUNITY test IEC 60601 test level...
2. Do not dispose of electric or electronic equipment together with unsorted municipal waste. website www.ameda.com 13. OPERATIONAL & STORAGE CONDITIONS 104°F 131°F...
If you will be traveling outside of your country where you received your Ameda Mya Joy breast pump you will need to use batteries or a Ameda Mya Joy AC power adapter specific to that area of travel in order to use your breast pump while traveling.
Page 25
Mya Joy To register your pump: Please visit ameda.com/register-your-pump and input the serial number (SN) from the bottom of your pump. Distributed in USA by: Distribution in other countries: For product information or Distribué aux États-Unis par : feedback, call your local distributor or location where you Distribuido en los EE.
Need help?
Do you have a question about the MyaJoy and is the answer not in the manual?
Questions and answers