Ameda MyaJoyPlus Instructions For Use Manual

Double electric breast pump
Hide thumbs Also See for MyaJoyPlus:

Advertisement

Double Electric Breast Pump
EN
Instructions for use
Ameda.com || 866.992.6332 || ParentCare@ameda.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the MyaJoyPlus and is the answer not in the manual?

Questions and answers

Summary of Contents for Ameda MyaJoyPlus

  • Page 1 Double Electric Breast Pump Instructions for use Ameda.com || 866.992.6332 || ParentCare@ameda.com...
  • Page 2 The Ameda HygieniKit is a closed system that contains a diaphragm that keeps the air from the pump and tubing separate from the pump parts that are exposed to your breast milk, so your tubing stays dry. This diaphragm helps prevent milk backup into tubing.
  • Page 3: Table Of Contents

    AMEDA ACCESSORIES AND SPARE PARTS ........
  • Page 4: Warnings & Cautions

    • Prior to use, always inspect power adapter for damage or exposed wires. If damage and/or exposed wires are found, do not use power adapter and call Ameda Customer Service ParentCare at 1.866.992.6332. • NEVER operate electrical product: if it has a damaged cord or plug; if it is not working properly; if it has been dropped, damaged, or becomes wet.
  • Page 5 • DO NOT wrap cord around the body of the adapter. • Use ONLY the AC power adapter that is recommended for the Ameda Mya Joy PLUS Breast Pump. • DO NOT turn on the breast pump in an aircraft.
  • Page 6: Intended Use

    SAVE THESE INSTRUCTIONS 2. INTENDED USE The Ameda Mya Joy PLUS Breast Pump is a single-user, powered breast pump intended to express and collect milk from the breasts of lactating women. 3. CONTRAINDICATIONS FOR USE There are no known contraindications for use with this product.
  • Page 7: Product Description

    4. PRODUCT DESCRIPTION Ameda Mya Joy PLUS & HygieniKit Milk Collection System ® Pump Assembly 1. Ameda Mya Joy PLUS Pump Unit 2. AC Power Adapter (with USB-C connection) 3. Tubing Adapter 4. Tubing 5. Adapter Cap 6. Diaphragm Ameda HygieniKit 7.
  • Page 8: Set-Up Instructions

    CLEAN TUBING: It is necessary to wash the tubing only if condensation is present. If condensation is present in the tube, wash and dry it immediately or replace the tubing. If milk is present in the tubing, do not attempt to wash or clean the tubing. Instead contact Ameda Customer Service at 1.866.992.6332. HOW TO WASH TUBING: Rinse the tubing by pouring cool water and make sure the water flows through the entire tubing from one end to the other end.
  • Page 9: Assembling The Hygienikit

    5.2 Assembling the HygieniKit Milk Collection System ® Once your pump parts are completely dry before use, wash your hands thoroughly and assemble the HygieniKit ® Milk Collection System (Pump Kit). Adapter Cap Diaphragm Breast Flange Tubing Valve Tubing Adapter Bottle HOLD by the rim or thicker PUSH the valve onto the...
  • Page 10 ATTACH and TIGHTEN the bottle or INSERT the diaphragm into the top attach milk storage bag. of the flange. It is important that the inside of the diaphragm is completely DRY to avoid moisture from forming in the tubing. ATTACH the adapter cap(s) by The tubing should still be CONNECTED pressing it onto the top of the to the adapter cap(s) and tubing adapter.
  • Page 11: Inspecting Before Pumping

    Choosing and checking your electrical source AC POWER ADAPTER WARNING: Use ONLY the AC power adapter recommended for the Ameda Mya Joy PLUS Pump. WARNING: Always inspect the AC power adapter for damage such as kinking or twisting of cords or frayed or exposed wires.
  • Page 12: Lcd Icons & Modes

    NOTE: After two minutes, the pump will automatically switch to Expression Mode unless you manually adjust beforehand. C. SUCTION STRENGTH LEVEL Level of suction strength during pumping. The Ameda Mya Joy PLUS has two suction modes (Dual-Phase), each with adjustable suction levels: – Stimulation Mode: 1 - 6 levels –...
  • Page 13: To Begin Pumping

    7.4 To Begin Pumping STEP 1: INSERT TUBING Insert Tubing adapter into the right side of the pump base, then turn the pump slightly to the right, and gently push it to make sure it connects firmly. A. For DOUBLE PUMPING: Both tubes should be attached to the tubing adapter.
  • Page 14 STEP 4: ADJUST VACUUM LEVEL Adjust the Stimulation Vacuum Level between 1 and 6 to your preferred comfort level. Never set to uncomfortable level. TIP: Start off with the suction low, and then gradually increase the suction setting to your most comfortable level. You may find that the level of suction that you are comfortable with varies from day to day, during the day and even during the same expression session.
  • Page 15: How To Fit Your Flange

    7.5 How to Fit Your Flange Check breast flange fit, if pumping causes discomfort a larger or smaller breast flange size may be needed. GOOD FIT FLANGE TOO SMALL FLANGE TOO LARGE During pumping, your nipple During pumping, some or all During pumping, more areola moves freely in the breast flange of your nipple rubs against the...
  • Page 16: Cleaning The Hygienikit

    8. CLEANING THE HYGIENIKIT ® BEFORE EVERY USE Wash hands with soap and water. Inspect and assemble clean pump kit. If your tubing is moldy, discard and replace immediately. Clean pump dials, power switch, and countertop with disinfectant wipes, especially if using a shared pump. AFTER EVERY USE Store milk safely.
  • Page 17: Maintenance

    9. MAINTENANCE The breast pump base does not require service or maintenance. 9.1 Breast Pump Valve and Diaphragm Check Inspect the valves and diaphragms for tears or holes. Replace any damaged valves and diaphragms prior to pumping. Wash valves and diaphragms per cleaning section. Do not insert bottle brush or something similar into valves.
  • Page 18 (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. Caution: Changes or modifications to this device not expressly approved by Ameda may void the user's authority to use the device.
  • Page 19: Warranty

    90 days on other pump components except for breast pump valves. Ameda’s sole obligation under this limited warranty shall be to repair or replace, at Ameda’s option, any Product that is determined to be defective by Ameda and determined to be covered by this express limited warranty. Repair or replacement under this limited warranty is the sole and exclusive remedy of the First User.
  • Page 20: Technical Data

    12. TECHNICAL DATA 12.1 Specifications Product Description Electric Breast Pump Model Ameda Mya Joy PLUS The pump is capable of single and double pumping. Suction Stimulation mode: Dual-Kit: 40 to 100 mmHg Stimulation mode: Single-Kit: 80 to 160 mmHg Expression mode: Dual-Kit: 50 to 250 mmHg...
  • Page 21: Electromagnetic Compatibility

    12.2 Electromagnetic Compatibility The essential performance of the Mya Joy PLUS breast pump is to not exceed 270 mmHg. The pump was tested for immunity to electromagnetic disturbances and passed according to these criteria. TABLE 1 Guidance and declaration of manufacturer-electromagnetic emissions. The device is intended for use in the electromagnetic environment specified below.
  • Page 22 TABLE 2 Guidance and declaration of manufacturer-electromagnetic immunity. (For home healthcare environment) The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that it is used in such an environment Electromagnetic environment - IMMUNITY test IEC 60601 test level...
  • Page 23 TABLE 3 Guidance and declaration of manufacturer-electromagnetic immunity. (For home healthcare environment) The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that it is used in such an environment. Electromagnetic environment IMMUNITY test IEC 60601 test level...
  • Page 24 TABLE 4 IEC 60601-1-2:2014 Table 9 — Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment Test Maximum IMMUNITY Band Distance frequency Service Modulation power TEST LEVEL (MHz) (MHz) (V/m) Pulse 380 - TETRA 400 modulation 18 Hz 430 - GMRS 460, 5 kHz deviation...
  • Page 25: Product Disposal

    3. For further information regarding the environmental performance of this product please visit our website www.ameda.com. The breast pump and parts are made of plastics which have an impact on the environment when disposed of. Please recycle according to local regulations.
  • Page 26: Ameda Accessories And Spare Parts

    15. AMEDA ACCESSORIES AND SPARE PARTS SOME PUMPS MAY ALSO INCLUDE SOME OF THESE ACCESSORIES: You may also be able to order these additional accessories or spare parts online, at a local Ameda retailer or Ameda insurance covered provider, visit www.ameda.com for more information.
  • Page 27: Symbols Reference

    16. SYMBOLS REFERENCE ISO 15223-1, Medical devices – Symbols to be ISO 15223-1, Medical devices – Symbols to be used with medical device labels, labelling and used with medical device labels, labelling and information to be supplied, Part 1: General information to be supplied, Part 1: General requirements,Clause 5.3.1, Fragile, handle with requirements, Clause 5.1.7 Serial number /...
  • Page 28 To register your pump: Please visit ameda.com/register-your-pump and input the serial number (SN) from the bottom of your pump. Distributed in USA by: Ameda, Inc. 485 Half Day Road Buffalo Grove, IL 60089 1.866.992.6332 www.ameda.com Distributed in Canada by: Mothers Choice Products Ltd.

Table of Contents