USER GUIDE: FILLMASTER 3400 & 3600

CALIBRATION

Checking Calibration:
Press "START" on the dispenser.
The display will read "SELECT KEY> ENTER AMOUNT___. "
Type in 50ml, then press "ENTER. "
Place a 100ml graduate under the dispensing tip.
Press "DISPENSE" to begin water flow.
If the amount dispensed is 50ml, your Fillmaster is properly calibrated.
If the amount dispensed is less or more than 50ml, your Fillmaster must be
calibrated.