USER GUIDE: FILLMASTER 3400 & 3600

WATER DISPENSING

Press the "Start" button
The display will read "SELECT KEY> ENTER AMOUNT___. "
Type in the desired water volume in ml, then press "ENTER. "
Place a graduated cylinder under the dispensing tip (be sure the graduate you are
using is of sufficient volume to hold the amount of water to be dispensed).
Press "DISPENSE" to begin water flow
For volumes higher than 30ml, the unit will dispense water in two equal
parts. Always remember to dispense the second half of the water.
Check the graduated cylinder to ensure the desired amount of water was
dispensed.
If the desired amount of water is not dispensed, your dispenser may need to
be calibrated or the battery may need to be changed. See the "Calibration"
and "Power Supply" sections below for instructions on how to check
calibration or battery life.
Water volumes of 13ml or lower may appear to under or over dispense, even if
your dispenser is properly calibrated. This is normal and does not necessarily
mean your dispenser is out of calibration or defective.
If the machine is under dispensing at a low volume, use the unit to
add water as needed.
If the unit is over dispensing at a low volume, pour out the amount
that was over dispensed.
Pour water from graduate into medicine bottle to complete the process.
For volumes of water higher than 30ml, dispense the water into the
medication bottle in two equal parts to allow for mixing.
For volumes of water less than or equal to 30ml, dispense the entire amount
into the medication bottle and mix.
Fillmaster Systems, LLC | 9475 Gerwig Lane, Columbia MD 21046
Fillmaster dispensers are designed to quickly and
accurately dispense water for reconstitution of
medications. Should you encounter any issues
with your Fillmaster dispenser, consult the
Troubleshooting guide or call 888-284-7873 for
assistance.
Phone: 888-284-7873 | www.f i llmastersystems.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 3400 and is the answer not in the manual?

Questions and answers

Summary of Contents for Fillmaster Systems 3400

  • Page 1 • For volumes of water less than or equal to 30ml, dispense the entire amount into the medication bottle and mix. Fillmaster Systems, LLC | 9475 Gerwig Lane, Columbia MD 21046 Phone: 888-284-7873 | www.f i llmastersystems.com...
  • Page 2 The battery life of the iPOWER Pak can be checked by pressing the “Program” button 3x, and selecting option “2” for Power Supply. A gas gauge style power level will be displayed on the screen. Fillmaster Systems, LLC | 9475 Gerwig Lane, Columbia MD 21046 Phone: 888-284-7873 | www.f i llmastersystems.com...
  • Page 3 If the amount dispensed is less or more than 50ml, your Fillmaster must be calibrated. Fillmaster Systems, LLC | 9475 Gerwig Lane, Columbia MD 21046 Fillmaster Systems, LLC | 2030 Gillespie Way, Suite. 104, El Cajon, CA 92020 Phone: 888-284-7873 | www.f i llmastersystems.com Phone: 888-284-7873 | www.fillmastersystems.com...

This manual is also suitable for:

3600

Table of Contents