Technical Data; Control Unit - Agilent Technologies 729-0800 Instruction Manual

Ion pump battery power supply
Hide thumbs Also See for 729-0800:
Table of Contents

Advertisement

TECHNICAL DATA

Control unit

Power supply:
a commercially available 9V PP3 type non-
rechargeable battery (6AM6, MN1604, 6LR61)
(recommended type Duracell Plus).
an input connector for connection and a 9V AC/DC
converter (+/- 10%)
The negative of the 9V power supply is on the
central pin
The positive of the 9V power supply is on the
external connection.
Output:
Non regulated voltage of between 3.0 and 1.5 kV
for 829-XXXX and from –3,0 to –1,5 kV for
729-XXXX according to battery charge level
Factory-set negative output polarity
maximum output current 1 µA (corresponding to 5
and - 9 mbar in a 55/LS ion pump)
Max. output power: 3 mW
Control panel:
Power ON/OFF switch
Two LEDs: L1 green
L2 red
The meaning of the configurations of the LEDs is
shown in the table.
L1
HV ON
OFF
PUMP ON
HV ON
BLINKING
PUMP OFF
BATTERY
OFF
DISCHARGED
The two LEDs blink at a frequency of 2 seconds
(1.8 second off, 0.2 seconds on).
WARNING
L2
BLINKING
OFF
OFF
Panel
Life:
Minimum 30 days of operation with an output
current of 1 µA
Connection:
HV cable connected internally
Dimensions:
150x80x55 mm
Environmental conditions:
Storage temperature: from –40 to +85 °C
Operating temperature equal to the operating
temperature of the battery
Relative humidity: from 10% to 90%
The unit is furnished with the negative output
polarity and type of connector requested.
The unit may be supplied with three different pump
connectors.
2
4
1.
Power supply unit
2.
Pump power connector
3.
Ground terminal
4.
External voltage input connector
5.
ON/OFF switch
6.
Operating state readout led
14/16
TECHNICAL INFORMATION
1
3
5
6
87-900-111-01 (C)

Advertisement

Table of Contents
loading

This manual is also suitable for:

729-0801729-0802829-0800829-0801829-0802

Table of Contents