Entering Calibration Data into an HP 8702A/B

Before using your source or receiver with an HP 8702A/B lightwave compo-
nent analyzer, you must load the calibration data as described in this section.
Two procedures are provided in this section. The first procedure is the stan-
dard method. You will most likely use it. However, if you have an HP-IB exter-
nal disk drive, you can use the second procedure. For more information on
calibration data,
refer to "Entering Calibration Data into an HP 8702D" on
page
1-9.

To install from the front panel

The following procedure is written specifically for entering source data. If you
are entering data for a receiver, simply substitute the word RECEIVER for SOURCE
and the word RCVR for SRC .
1 On the HP 8702A/B, press:
PRESET
CAL , CAL KITS & STDS , SOURCE [COEFF] , SRC COEFF , ENTER SRC COEFF
2 Use the available softkeys to enter the coefficients that are listed on the label
that is attached to the source.
3 Press LABEL STD , and enter a label for the calibration data.
4 Press STD DONE and then SAVE SRC COEFF .