Getting Started
Entering Calibration Data into an HP 8702D
To copy the data from the disk
The following procedure is written specifically for entering source data. If you
are entering data for a receiver, simply substitute the word RECEIVER for SOURCE
and the word RCVR for SRC .
1 Locate the DOS - formatted calibration disk provided with the source.
2 Place the disk in the HP 8702D's front-panel disk drive.
3 Press CAL , CAL KITS & STDS , SOURCE STANDARDS , LOAD SRC DISK MENU . This will
display the disk directory screen.
4 Use the arrow keys or front panel knob to select a cal data file.
5 Load the cal data file into either of two internal registers by pressing the
corresponding LOAD softkey.
To manually enter data from the label
The following procedure is written specifically for entering source data. If you
are entering data for a receiver, simply substitute the word RECEIVER for SOURCE
and the word RCVR for SRC .
1 Press CAL , CAL KITS & STDS , SOURCE , CAL STD:SRC COEFF and then ENTER SRC COEFF .
2 Press each softkey corresponding to the nine coefficients (A through I), using
the MORE softkey, and enter the coefficients. The coefficients are listed on a
label that is on the source. After all of the coefficients are entered, press
RETURN .
3 Press LABEL STD .
4 Use the front-panel knob and the SELECT LETTER softkey to write a title for the
calibration data.
5 Press STD DONE (DEFINED) and then SAVE SRC COEFF .
6 Press CAL , CALIBRATE MENU , RESPONSE , and then SOURCE .
7 When the SOURCE softkey is underlined, press DONE: RESPONSE .
1-10