Endress+Hauser Nivotester FTC325 Operating Instructions Manual page 23

Level limit switch with 3-wire technic
Hide thumbs Also See for Nivotester FTC325:
Table of Contents

Advertisement

12 pF
27 pF
45 pF
Setting-up
Example:
Conductivity Build-up
εr
Material
Solvents,
< 3
low
fuels
Dry
< 3
low
bulk solids
Moist
> 3
average
bulk solids
Aqueous
> 3
strong
liquids and
alcohols
Sludge
> 3
strong
Endress+Hauser
0 pF
0 pF
50 pF
Type of
probe mounting
Insulation
Ground
tube
Full Part
with without
low
low
average
low
strong
very
strong
C
25 pF
Standard
switch position
Standard
Operation
operation
as
overspill
protection
16 pF
16 pF
16 pF
25 pF
25 pF
22 pF
36 pF
28 pF
50 pF
en - Switch point offset for
build-up compensation
Switch position and switch point
offset at initial capacity
(CA) = 40 pF
0.8 pF= max. sensitivity
50.0 pF= min. sensitivity
For the adjustment, the switch
must be turned to the leftmost
position.
Caution!
The device might no longer
switch if the switch point offset
is too big!
23

Advertisement

Table of Contents
loading

Table of Contents