Foreword Dear Ladies and Gentlemen, The stakpure GmbH team thanks you for the trust you have placed in us. By choosing this ultrapure water system from our company, you have opted for an innovative, high- quality and durable product. Before installing and starting up your ultrapure water system, please read the installation and operating instructions in this operating manual carefully.
Symbols used Warning notices with this symbol indicate a hazard which, if not avoided, could result in death or serious injury. Warning notices with this symbol indicate a hazard which, if not avoided, could result in minor or moderate injury. This symbol indicates a risk with possible property damage.
2 Safety instructions The safety instructions described below are for your own safety and help to prevent possible damage to the OmniaTap xs system. Read the instructions before installation, commissioning and maintenance and follow them carefully. Danger of electric shock! Improper electrical supply of the OmniaTap xs system can lead to an electric shock! For the electrical supply of the OmniaTap xs system, only use the wide-range power pack included in the scope of delivery.
Page 9
Danger of skin and eye injuries! Contact with the disinfectant can lead to skin and eye irritation and/or injuries! Always wear appropriate protective clothing (at least gloves & goggles) when disinfecting the OmniaTap xs system to avoid contact with the disinfectant. Make sure that no disinfectant can escape uncontrolled from the OmniaTap system, check the connection hoses for correct and tight fit.
Depending on the country, a safety combination consisting of a backflow preventer and a system separator may have to be available on the building side. A DN 50 floor drain with free drainage is to be provided in the operating room. If there is no floor drain, we recommend using a water monitor to prevent damage to the system and equipment as well as the building.
4 Transport and packaging The OmniaTap xs systems are carefully inspected and packaged prior to shipment, however damage may occur in transit. Check the packaging and the system for possible transport damage. If you find any damage, please contact the shipping or transport company responsible for the delivery directly. For transport, only lift the device at holding points 1, 2 and 3.
5 Delivery package The scope of delivery of the OmniaTap xs system consists of the following parts: Position Number Designation touch OmniaTap xs Blueline System Table power supply 48V 120W Power cord PE hose d8 2,5m PE hose d8 incl. input protection filter and Water connection adapter R ¾“...
Page 13
Product water ASTM II touch touch OmniaTab xs Blueline OmniaTap xs Blueline 8 UV-TOC 8 UV-TOC/UF Conductivity [µS/cm] at 25°C 0.067 – 0.1 0.067 – 0.1 Resistance [MΩ cm] at 25°C 15 – 10 15 – 10 Retention rate for germs and particles 99 % 99 % Permeate output / l/h at 15°C...
Page 14
Dimensions & Weight touch touch OmniaTab xs Blueline OmniaTap xs Blueline 8 UV-TOC 8 UV-TOC/UF Height [mm] Depth [mm] Width [mm] Tank capacity (useful volume) [l] Empty weight kg] Operating weight [kg] Materials of the components in contact with water Solenoid valve input POM, EPDM Flush solenoid valve...
7 Flow charts touch 7.1 Flow chart OmniaTap xs UV-TOC touch 7.2 Flow chart OmniaTap xs UV-TOC/UF _________________________________________________________________________________ touch OmniaTap xs ; Version November 2022 Subject to technical changes...
Page 16
Designation Designation Pretreatment cartridge QIA conductivity cell TOC Ultrapure water cartridge TIA temperature sensor TOC Sterile filter Pure water tank Sterile ventilation filter Ultrafiltration module Hat filter (in the connection hose) UV Photooxidation Flow meter Inlet solenoid valve Level sensor Rinsing solenoid valve LSZ1 Leakage sensor...
8 System description 8.1 System structure 1: Detachable OptiFill dispenser with integrated 7: Ultrapure water cartridge control unit and USB port 8: UV reactor 2: Rotating dispenser arm 9: Ultrafiltration module 3: Snap closure 10: Dispenser admission 4: Removable cover hood 11: USB port 5: Pure water tank 12: Tank opening with lid...
8.2 System connections on the back Inlet: Feed water connection Drain: Flushing water connection Power: Power supply connection RS232: Connection for data printer (Option) 9 Function description The systems of the OmniaTap xs series prepare drinking water directly into pure and ultrapure water by using several treatment technologies.
10 Assembly 10.1 Operating environment When selecting the installation location, observe the following requirements: The pressure of the feed water must be between 0.5 and 6 bar. The standing surface must be level. A suitable protective contact socket must be available for the electrical supply of the OmniaTab xs system (see technical data).
Page 20
Water can escape through leaks. Therefore, always make sure that the floor in the area of the work surface is dry to avoid accidents caused by slipping. Use the connection accessories to connect the OmniaTap xs system to the feed water connection and to the free drain.
Page 21
Then screw the sterile filter with its G1/4" thread into the outlet of the OptiFill dispenser (see figure). _________________________________________________________________________________ touch OmniaTap xs ; Version November 2022 Subject to technical changes...
10.3 Wall mounting Risk of injury from falling and crushing! The OmniaTap xs system may only be attached to a suitable wall using the optionally available wall holder. Use the wall mount exclusively for the OmniaTap xs system. Make sure that the wall and the fastening material have a sufficient load-bearing capacity of at least 100kg.
Page 23
The next step is to attach the wall holder rail to the wall. This requires 6 holes with a diameter of 8mm and a depth of at least 60mm. Drill the holes according to the pattern shown in the "Wall Mount Rail" figure below. Pay attention to a horizontal alignment. Wall mount rail Then insert the dowels into the drilled holes and fasten the wall mounting rail to the wall with the screws so that the two mounting holes are facing upwards.
10.4 Installation of an optional pump station An optional pump station, article number 16580000, is available to supply an external consumer such as an analyzer directly with pure water from the tank of the OmniaTap xs system. The pump station can be connected directly to the tank of the OmniaTap xs system.
11 Operating concept The Omnia devices are operated using the three buttons below the display. If entries are required in the respective menus, an ESC and Enter key also appear as touch keys on the screen. You get to the menus by "swiping", if you push the main display up, you get to the customer menu. All other conductivities including limit value and temperature of the corresponding measuring cell, operating hours (pumps, UV) are displayed there.
11.2 Navigation Setup menu Calibrating menu Main display System- Consumption Consumption limits display settings _________________________________________________________________________________ touch OmniaTap xs ; Version November 2022 Subject to technical changes...
11.3 Operating concept After pressing the "Change" soft key, the selected line is highlighted in red. Another value can be selected by pressing the arrow key. The input is activated with the drop key and the cursor jumps to the first digit. Use the drop button to move the cursor to the position to be changed and use the arrow keys to set the desired value.
11.4 Volume metered water to be withdrawed The withdrawal mode is activated by pressing the drop button, recognisable by a drop indicator under the withdrawal tap on the display. In the withdrawal display you have the option to select one of the preset withdrawal values.
11.6 Setup menu Consumption Limits Limit value Limit value Limit value TOC Monitor Limit value Limit value Pump Pump Circulation exchange Rinsing Date&Time Change Emergency supply _________________________________________________________________________________ touch OmniaTap xs ; Version November 2022 Subject to technical changes...
12 Commissioning As soon as the power supply to the OmniaTap xs system is established, the system starts and the system name, serial number and software version are displayed for 3 seconds. The display then changes to the main display. Start Main display OmniaTap...
13 Maintenance Regular maintenance and care of the OmniaTap xs system is required to ensure consistently good water quality. For professional and regular maintenance of the OmniaTap xs system, we recommend concluding a maintenance contract with an authorised customer service. If improper maintenance, care and repair work is carried out on the system, the warranty for this system will expire.
13.2 Changing the cartridges The cartridges have to be changed at regular intervals (12 months) or if the limit value you have specified is exceeded for a long time. The start of the interval is the point in time at which the respective filter cartridge was first connected to the system.
13.3 Disinfection To protect the OmniaTap xs system from contamination from biological deposits, disinfection should be carried out at regular intervals (12 months). For the implementation you need the disinfection kit article no.: 19200091. Risk of chemical burns! The disinfection of the OmniaTap xs system may only be carried out by appropriately qualified specialist personnel.
14 Decommissioning & Dismantling Decommissioning and dismantling may only be carried out by authorised specialist personnel. If the OmniaTap xs system is to be decommissioned and dismantled, proceed as follows: Disconnect the system from the power supply by pulling out the mains plug Disconnect all hoses from the system connectors.
18 Malfunctions, causes & solutions 18.1 Faults table Fault Possible cause Remedy System does not start / Ensure the power supply Display does not function No or faulty power supply according to the specifications under "6 Technical data". Check the hose connection of System is not receiving feed the feed water supply.
18.2 Error messages Error messages are flashing and shown in red on the display. If there are several errors, they are displayed one after the other. After the error has been rectified, the display goes out after approx. 15 seconds. Error messages Possible cause Remedy...
Page 40
The limit for the operating Uptime filter time of the ultrapure water Insert a new filter cartridge cartridge has been reached Defect of the Please contact customer Measuring cell LF1 Conductivity measurement LF1 service Defect of the Please contact customer Measuring cell LF2 Conductivity measurement LF2 service...
19 Disposal Stakpure ultrapure water systems are marked with a "crossed out trash can". This means that according to the European Regulation 2012/19/EU, your old electronic device may not be disposed of with normal household waste. You can hand in your...
20 EU declaration of conformity _________________________________________________________________________________ touch OmniaTap xs ; Version November 2022 Subject to technical changes...
Page 43
_________________________________________________________________________________ touch OmniaTap xs ; Version November 2022 Subject to technical changes...
Page 44
GmbH Auf dem Kesseling 11 D – 56414 Niederahr Ph.: +49 2602 10673-0 Fax: +49 2602 10673-200 info@stakpure.com www.stakpure.de 50003537...