Dispenser DP 30 Instructions for use 1. Product description Intended use and operation The Dispenser DP 30 serves as Infiltration pump into the connective tissue and is used in the following field of appli- cations: Tumescent infiltration for liposuction (Liposuction) ...
The use of (RF) Radio Frequency emitting devices and equipment as well as the occurrence of negative environmen- tal factors in the close area of the Dispenser DP 30 may cause unexpected or adverse operation. The connection or the placing of other devices in close vicinity is not allowed.
Repairs may only be performed by authorized Nouvag service technicians! The Dispenser DP 30 may only be operated under constant supervision of medical personnel. The absence of a warning buzz to indicate malfunctions of the device requires the permanent control of the volumetric displace- ment of the pump.
Dispenser DP 30 control unit 1770 Stand for irrigation fluid bottle 31678 Dispenser DP 30 Instructions for use Selectively – Set No. 4186, Dispenser DP 30 control unit with ON/OFF pedal Description Quantity 1513nou ON/OFF pedal, IPX8 Selectively – Set No. 4187, Dispenser DP 30 control unit with VARIO pedal...
Beim Einlegen des Schlauchsets ist die Pfeilmarkierung auf dem Schwenkarm der Pumpe zu beachten. Sie gibt die Flussrichtung der Infiltrations-Flüssigkeit an. Es darf nur das Nouvag Schlauchset verwendet werden, ansonsten kann die Funktion nicht gewährleistet werden. Bedienungsanleitung, Dispenser DP 30, Ref. 31678, V24/21...
Der Beutel der Infiltrationsflüssigkeit darf maximal 2 kg wiegen. Schwerere Beutel als erlaubt, Do not regulate the amount of infiltration fluid using the roller clamp on the tube set; with the Dispenser DP 30, Der Beutel der Infiltrationsflüssigkeit darf maximal 2 kg wiegen. Schwerere Beutel als erlaubt, können zum Umkippen des Geräts führen.
Wird während des Betriebes der Pumpe die Pumpen-Entriegelungstaste betätigt, stoppt die Pumpe. Functional check Prior to Dispenser DP 30 startup or use of accessory equipment, the user must always ensure that each individual Funktionskontrolle component is in good working order, free of defects, clean, sterile and operational. The tube set has to correspond Vor jeder Inbetriebnahme des Dispenser DP 30 und dessen Zubehör hat sich der Anwender vom ordnungs-...
The STI (Safety Technical Inspection) for the Dispenser DP 30 shall be executed every 2 years by authorized experts. Die STK für den Dispenser DP 30 ist alle 2 Jahre und nur durch autorisierte Stellen durchzuführen und zu Results shall be documented.
Page 16
Electromagnetic compatibility (EMC) Remark: The Product subsequently referred to herein always denotes the Dispenser DP 30. Changes or modifications to this product not expressly approved by the manufacturer may result in increased emissions or decreased immunity performance of the product and could cause EMC issues with this or other equipment. This product is designed and tested to comply with applicable regulations regarding EMC and shall be installed and put into service according to the EMC information stated as follows.