CONTENT I. Safety rules and precautions .................. 3 II. Introduction ....................... 3 III. Range and accuracy ....................3 IV.Model of Series ....................4 V.Technical Specification ..................4 VII.Operation ......................5 - 2 -...
0.001mΩ~ 300.0KΩ. The upper computer software has functions such as historical data reading, consulting, saving, and reporting. III. Range and accuracy 0.01mΩ~100.00mΩ 0.01mΩ 1.2A 100.1mΩ~1000.0mΩ 0.0001Ω 1.2A 0.001Ω 1.2A 1.001Ω~10.000Ω Within 18℃~28℃, 5004 10.01Ω~100.00Ω Below 70%rh: 0.01Ω 0.5A ±0.1%FS±20dgt 100.1Ω~1000.0Ω 0.0001KΩ 0.05A 0.001KΩ 1.001KΩ~10.000KΩ 0.01KΩ 0.5mA 10.01KΩ~30.00KΩ...
IV . Model of Series Resistance Range 0.01mΩ~30.00KΩ 0.001mΩ~300.0KΩ 10uΩ 1uΩ Resolution ±0.1%FS ±0.1%FS Accuracy Data storage 500 sets 500 sets USB Data upload function Charging function V . Technical Specification Mainly used to measure the wire resistance of cables, the contact resistance of switches, connectors, relays, the resistance of coils, motors, transformer winding and metal riveting Function resistance, the connection resistance test between metal components, the low-value...
humidity Overload AC 220V/0.0001 second between C1-C2 P1-P2 ports, after overload protection, you need to protection restart the meter to test normally Insulation More than 10MΩ(Between circuit and case 500V) resistance Pressure resistance AC 3700V/rms(Between circuit and case) Electromagnetic IEC61010-4-3,Wireless frequency electromagnetic field≤1V/m properties IEC61010-1, CAT III 600V, Pollution Level 2, JJG724-1991 "Determination of Direct Current Suitable for Safety...
Page 7
The connection between the test line, the detector and the object under test must be reliable. Since component test heating will cause errors, it is recommended that the test time should not exceed 30 seconds, and the interval between each test should be 30 seconds, the measurement results will be more accurate.
Page 8
The following figure shows an example to test the resistance value of the contact point of the relay. 5. Backlight control After booting, press the " " key to turn on or off the backlight. The backlight function is suitable for dim places. The default backlight is off every time you turn on 6.
Page 9
7. Data access/delete After turning on or after the measurement, press the " " key to switch to the data access mode, and the word "MR" is displayed below. Press the “ ” or “ ” key to set the step value to the corresponding data of the storage data group of 1 group, and press the “...
Page 10
10. Battery description Generally charge for 5 to 8 hours. If the detector is left unused for a long time, please charge the battery every 1 to 2 months. Please use the original charger for charging. This machine does not support fast charging.