Stirring Fluid Loss Manual Introduction The Stirring Fluid Loss Assembly simulates filtration at downhole conditions in a wellbore. This test instrument determines the volume of fluid loss (filtrate) from a cement slurry, water-based mud or fracturing fluid at a set temperature and pressure. The test results are used to evaluate the fluid-loss control properties.
Stirring Fluid Loss Manual Document Conventions The following icons are used in this manual as necesary. NOTE. Notes emphasize additional information that may be useful to the reader. CAUTION. Describes a situation or practice that requires operator awareness or action in order to avoid undesirable consequences. MANDATORY ACTION.
Stirring Fluid Loss Manual Safety Safe operation of the Stirring Fluid Loss Test Assembly requires the operator understand and practice the correct assembly and operation of the equipment. Improper assembly, operation, or the use of defective parts increases the likelihood of cell leakage or failure, which could result in serious injury and damage.
Stirring Fluid Loss Manual Safe Electrical Operation Make sure the electrical source is fused and grounded. Verify the power cord on the instrument is in good condition and that it has the proper ground connection. Electrical problems may not be obvious by looking at the equipment. If the unit blows fuses or trips breakers, the heating time seems longer than normal or the temperature controller does not reliably maintain temperature, then electrical repair may be required.
Stirring Fluid Loss Manual Features and Specifications The Stirring Fluid Loss Test Assembly is equipped with a stirrer to simulate circulation and a heating jacket to simulate the bottomhole circulating temperature (BHCT). The pressure differential between annular and formation pressure is simulated by pressurized nitrogen.
Stirring Fluid Loss Manual Table 3-1 Stirring Fluid Loss Test Assembly Specifications Category Specification Maximum Temperature 400°F (204°C) Maximum Pressure 2,000 psi (13.8 MPa) Pressure Required 1,500 psi (10.3 MPa), Nitrogen gas Filtering Area 3.5 in (22.6 cm Sample Chamber (1) 3.25 x 12.0 in Diameter x Length (excludes valves)
Stirring Fluid Loss Manual Figure 3-2 Paddle Shaft and Cell Cap 3.1.2 Sample Chamber The sample chamber (cell or reservoir) is where the slurry sample is added and stirred at set temperature and time (conditioning). The paddle cap assembly is inserted into the sample chamber.
Stirring Fluid Loss Manual The back pressure unit is necessary when testing cement slurry containing salt at test temperatures greater than 212 F (100 If back pressure is not applied, the filtrate will flash to steam when exiting the fluid loss cell and the salt will plug the small capillary tube that goes through the standard filtrate condenser.
Stirring Fluid Loss Manual 3.2.1 Electrical Controls The electrical controls and their functions are described below. • Master Power Switch (Master) controls all the electrical power to the different controls. • Temperature Controller (Temperature Control) controls the rate of the temperature rise (ramp) and the final set point temperature of the test slurry.
Stirring Fluid Loss Manual 3.2.2 Pressure and Coolant Controls Fluid loss tests performed with the provided reservoir should not be performed with pressures exceeding 1,500 psi (6.9 MPa). The pressure and coolant controls and their features are described below. Figure 3-7 Pressure and Coolant Controls •...
Page 18
Stirring Fluid Loss Manual o Air Cool for air flow; arrow points to the left o Off to shut off; arrow points up 100031980 Revision F...
There should be sufficient storage area for commonly used tools and consumable items. Customer installed wiring, electrical connectors, and power cords are excluded from the warranty. A Fann representative is available to install and set-up the instrument if necessary. Figure 4-1 Back Panel 100031980 Revision F...
Stirring Fluid Loss Manual Operation This section gives detailed instructions on operating the instrument. The following sections describe how to • Pre-test the Instrument • Load the Sample Chamber • Pre-heat the Sample Chamber • Start the Stirring Fluid Loss Test •...
Stirring Fluid Loss Manual Figure 5-1 Flexible Drive Shaft 14. Turn the instrument on by moving the Master switch (front panel) to the On position. 15. Start stirring motor by moving the Motor switch (front panel) to the On position. Note that the paddle and shaft are turning properly. If necessary, review Section 7 for troubleshooting and maintenance instructions.
Stirring Fluid Loss Manual 26. Check for leaks. If there are no leaks, bleed off. If there are leaks, then check the valves, O-rings, packing, and make adjustments. 27. Remove the cell cap, O-rings and brass ring and proceed with adding the sample and running the test.
Stirring Fluid Loss Manual Figure 5-3 Screen and O-ring in Sample Chamber 6. Screw cell cap onto end of reservoir. It should be as nearly flush as possible with reservoir end. Use special wrench to tighten the cap. 7. Close the supply pressure valve on the cap. 8.
Stirring Fluid Loss Manual Figure 5-4 Thermocouple inserted in sample cell 3. Turn on the Heater. Eurotherm EPC3004 Display • Hold the Page button (~ six seconds) until Level 1 appears then use to Level 2 press Scroll • Press Page until (Program Setup) •...
Stirring Fluid Loss Manual Start the Stirring Fluid Loss Test 1. Continue stirring sample until it reaches BHCT (bottomhole circulating temperature) and maintains that temperature for 30 minutes. 2. At 30 minutes after BHCT is reached, connect filtrate unit to sample chamber. Hand- tighten the filtrate unit.
Stirring Fluid Loss Manual Figure 5-6 Coolant Valve open for water flow 10. After sample reaches BHCT, open valve between sample chamber and filtrate unit. 11. Start Timer. Begin timing fluid loss volume (mL) when valve is opened. If slurry samples dehydrate completely (blow out) in less than 30 minutes, you will hear a hissing sound (gas released).
Stirring Fluid Loss Manual Figure 5-7 Coolant Valve in off position 5. Disconnect coolant tubes from filtrate unit and reconnect to heating jacket. C – 121 6. When temperature controller display indicates 200 F - 250 F (93 rotate the Coolant valve slowly to flow water through the heating jacket cooling coils.
Stirring Fluid Loss Manual 8. Turn nitrogen supply and coolant valves off. 9. Release pressure setting on regulator and open nitrogen vent valve. 10. Close the sample chamber pressure supply valve. 11. Disconnect sample chamber pressure line. Carefully remove sample chamber from heating jacket. Sample chamber may still contain pressure.
Page 29
Stirring Fluid Loss Manual 6. Pull the red release knob to unlock the heating jacket and rotate the heating jacket. 7. Push the red release knob to latch the chamber again. 8. Put the screen end together and test the sample chamber for leaks. Test at 1000 psi.
Stirring Fluid Loss Manual Test Analysis References API Recommended Practice for Testing Well Cements, API RP 10B-2. Results Test results will vary. According to API RP 10B-2, fluid loss tests that run the full 30 minutes typically show 5% variability. However, potential error increases for tests that run less than 30 minutes.
Stirring Fluid Loss Manual When a condenser is used, the filtrate volume in the condenser should be recorded. Calculations • For tests that run the entire 30 minutes without blowing dry, calculate API Fluid Loss as follows: API Fluid Loss = 2 Q where Q is the volume of filtrate collected at 30 minutes in ml.
Stirring Fluid Loss Manual Troubleshooting and Maintenance Regular maintenance procedures are provided in the following sections. If your equipment needs repair or service, please contact your Fann representative. Preventive Maintenance Table 7-1 Preventive Maintenance Schedule Frequency of Part Name Required Accuracy...
Stirring Fluid Loss Manual 10. Check parts thoroughly for wear and erosion. 11. Reassemble parts in reverse order of disassembly. Temperature Controller and Heating Jacket Use this procedure to test the temperature controller and heating jacket. 1. Connect a water supply of at least 40 psi to the test unit. 2.
Stirring Fluid Loss Manual Pressure and Motor Check Follow these steps to ensure proper operation: 1. Rotate the heating jacket until the handle and lock clamp are up. Verify the pressure tubing does not interfere with the rotating mechanism or heating unit. 2.
Stirring Fluid Loss Manual 13. Disconnect the drive and flip the chamber. 14. Turn off the nitrogen supply valve. 15. Slowly open the nitrogen vent valve until the gauge reads 300 psi. 16. Turn the regulator handle out until gauge needle just moves and venting is barely heard and reads 200-250 psi on the gauge.
Stirring Fluid Loss Manual Troubleshooting Refer to Table 7-2 for assistance in troubleshooting problems with the instrument. Table 7-2 Troubleshooting Guide Problem or Symptom Possible Cause Corrective Action The power source is Check the power source. disconnected or turned off. The main power switch has Check or replace the main The controller does not...
Page 37
Stirring Fluid Loss Manual Problem or Symptom Possible Cause Corrective Action The temperature reading is Possible short circuit in Look for and repair the about room temperature, thermocouple or short in the thermocouple and the chamber is hot. thermocouple cable. connector.
Stirring Fluid Loss Manual Parts List Check the following tables and assembly drawings for any replacement parts and their respective part numbers. Table 8-1 Stirring Fluid Loss Assembly, 50/60 Hz, P/N 210194, Revision F Item Part No. Quantity Description 0002...
Stirring Fluid Loss Manual Table 8-9 Filtrate Chamber Assembly, Back Pressure, P/N 100020356, Revision B Item Part No. Quantity Description 0001 100020366 FILTRATE CHAMBER LID 0002 100020365 FILTRATE CHAMBER 0003 100020367 COOLING COIL VALVE, PLUG, 1/8 MNPT, BRASS, 3000 PSI, 250 F, BUNA N 0004 100000593 SEALS...
Page 56
Returns For your protection, items being returned must be carefully packed to prevent damage in shipment and insured against possible damage or loss. Fann will not be responsible for damage resulting from careless or insufficient packing. Before returning items for any reason, authorization must be obtained from Fann Instrument Company.