Page 1
Cement Curing Autoclave User Manual 100072577 Revision H...
Page 2
All rights reserved. No part of this work covered by the copyright hereon may be reproduced or copied in any form or by any means (graphic, electronic, or mechanical) without first receiving the written permission of Fann Instrument Company, Houston, Texas, USA.
5.8 Running a test ....................41 5.9 Ending the test ....................43 5.10 Remove the chamber ..................43 5.11 Cleaning the molds, pressure chamber, and autoclave ........44 Analyzing Results ....................45 Troubleshooting and Maintenance ..............46 100072577 Revision H...
Page 4
7.1 Eurotherm 2404 Temperature Calibration ............46 7.2 Troubleshooting Tables ..................46 Accessories ......................50 8.1 Hand Pressure Pump ..................50 Parts Lists ......................51 10 Warranty and Returns ..................66 10.1 Warranty ......................66 10.2 Return of Items ....................66 100072577 Revision H...
Page 5
Figure 5-15 Water expelled as lid tightened ..............37 Figure 5-16 Handle spring loaded pin retracted ............. 38 Figure 5-17 Handle and manifold on chamber ............... 38 Figure 5-18 Place chamber into autoclave, remove handle ........... 39 Figure 5-19 Attach valve and manifold ................40 100072577 Revision H...
It is not recommended because of the danger of severe burns if touched or accidentally dropped. Also use extreme caution when placing a hot chamber in water. Hot steam generated when the water hits the hot chamber can cause severe burns. 100072577 Revision H...
Volts. There are electrically active terminals inside the instrument when the power switch is turned off. Physically disconnect the power cord from the outlet and instrument before attempting any electrical or mechanical maintenance. Refer to the electrical schematic before performing any maintenance or troubleshooting. 100072577 Revision H...
69 pounds (31.4 kg). Placing the top surface of the Cement Curing Autoclave at 24–36 in (61–91 cm) from the floor is an appropriate height for most people to lift the cement curing chamber. 100072577 Revision H...
Pressures above 3,000 PSI (20.68 MPa) are automatically vented by the pressure relief valve, which functions as an over-pressure safety device. The chamber pressure is shown on a pressure gauge installed directly into the top of the curing chamber. 100072577 Revision H...
Autoclave. The two mold halves are aligned with dowel pins, and clamped together with threaded sleeves. Two flat head screws fasten the bottom plate to the two mold halves. This arrangement allows the mold to be removed without damaging the cured cement specimen, and for thorough clean-up. 100072577 Revision H...
(set point) to the temperature reading from the curing chamber thermocouple and adjusts the heating cycle to achieve and/or maintain the desired temperature. It continuously displays the current and set point temperatures. 100072577 Revision H...
Model 2404 temperature controller when it is configured for the Cement Curing Autoclave. It is rare that the majority of these settings will need to be changed. They are provided for reference should the temperature controller respond in an unexpected manner, and incorrect settings are suspected. 100072577 Revision H...
Page 18
Cement Curing Autoclave User Manual 100072577 Revision H...
Page 19
Derivative type dtYP Front panel Auto/Man button EnAb Front panel Run/Hold button EnAb Power feedback PwrF Feed forward type Fwd.t nonE Sensor break output Sb.tr Sb.OP Forced manual output BCD input function nonE Gain schedule enable GSch LC.HI 100072577 Revision H...
Page 20
Sensor break impedance Auto Press to Set Point (SP) Configuration Name Value Parameter Description to select to change Number of setpoints Remote tracking rm.tr Manual track m.tr Setpoint rate limit units rmP.U PSec Remote setpoint nonE configuration 100072577 Revision H...
Page 21
AL 4 Latching Ltch Blocking bLoc Press to Program (PROG) Configuration Name Value Parameter Description to select to change Programmer type PtYP Holdback HbAC Power fail recovery Pwr.F cont Starting setpoint of a Srvo To.PU program (Servo point) 100072577 Revision H...
Page 22
Identity LoG.i Function of input Func nonE Press to Comms 1 Module (HA) Configuration Name Value Parameter Description to select to change Identity of the module nonE installed Function of input Func nonE 100072577 Revision H...
Page 23
%PID demand signal giving UAL.H maximum output Minimum average power Out.L Maximum average power Out.H 100.0 Press to Module 2 (2A) Configuration Name Value Parameter Description to select to change Identify of module installed nonE Function Func nonE 100072577 Revision H...
Page 24
Calibration point rcAL nonE User calibration enable UCAL Low calibration point for Input 1 Pt1.L High calibration point for Input 1 Pt1.H 999.9 Offset Low for Input 1 OF1.L Offset High for Input 1 OF1.H 100072577 Revision H...
Page 25
Cement Curing Autoclave User Manual Press to EXIT Press to select YES UNIT WILL RESTART Press to select FuLL 100072577 Revision H...
Page 26
Segment type tYPE rmPr Holdback type Target setpoint Ramp rate rAtE Segment number SEG.n Segment type tYPE rmPr Holdback type bAnd Target setpoint Ramp rate rATe Segment number SEG.n Segment type tYPE End of program End.t dwEll 100072577 Revision H...
Page 27
Derivative Time in seconds Manual Reset (%) High cutback Low cutback Proportional Band Pid 2 Integral Time in seconds 1700 Pid2 Derivative Time in seconds Pid2 Manual Reset (%) Pid 2 rES2 High cutback Pid2 Hcb2 Low cutback Pid2 Lcb2 100072577 Revision H...
Page 28
-199.9 Range high rnG.h 999.9 Press to Output list (oP) Name Value Parameter Description to select to change Instrument units uniT ºF Decimal places in the display dEc.P nnn.n Range low rnG.L -199.9 Range high rnG.h 999.9 100072577 Revision H...
Page 29
Parameter Description to select to change Instrument units uniT ºF Decimal places in the display dEc.P nnn.n Range low rnG.L -199.9 rnG.h Range high 999.9 Press to Page to ACCS list ‘Pass’ is displayed momentarily. Edit Press 100072577 Revision H...
Page 30
Instrument units uniT ºF Decimal places in the display dEc.P nnn.n Range low rnG.L -199.9 rnG.h Range high 999.9 Press to Page to ACCS list ‘Pass’ is displayed momentarily. Press to OPEr Press to Home View 100072577 Revision H...
Water Expelled as Lid is Closed Either pressure source is connected to the open 1/8 NPT port on the needle valve. After reaching the desired pressure, the needle valve can be closed and the pressure source disconnected. 100072577 Revision H...
As the curing temperature increases during the curing process, an automatic pressure relief valve prevents the chamber internal pressure from exceeding 3,000 PSI. This spring loaded valve releases small amounts of the pressurizing fluid (water). Figure 3-9 Initial pressure (left), increased pressure from temperature rise (right) 100072577 Revision H...
NEMA 5-20P (115V) or 6–15P (230V) plug, depending upon the voltage ordered. Because of the diversity in types of electrical outlets throughout the world, it may be necessary to replace this plug. Alternatively, the Cement Curing Autoclave can be directly wired into an 100072577 Revision H...
Page 34
115 VAC, 15 A for 115 VAC Cement Curing Autoclaves 4.4.2.2 230 VAC, 10 A for 230 VAC Cement Curing Autoclaves 4.4.3 Connect the power cord to the Cement Curing Autoclave, and then connect the power cord to the power source. 100072577 Revision H...
Autoclave chamber. To prevent the cement slurry from seeping into the joints between the sides and base plates, apply a thin coat of grease to the bottom edges of the mold. Keep and use like numbered cement mold halves together. Figure 5-1 Pins align the mold halves 100072577 Revision H...
Cement Curing Autoclave User Manual Figure 5-3 Grease mold for base plate 5.2.2 Attach the base plate to the mold using two flat head machine screws. Figure 5-4 Tighten base plate to mold 100072577 Revision H...
Excess grease at the inside joints of the molds will decrease the actual volume of the cement cube and can result in inaccurate compressive strength tests. Figure 5-5 Grease mold interior 100072577 Revision H...
5.4.1 Pour the cement slurry into the assembled mold. Repeatedly puddle and stir the slurry, then remove the excess with a straight edge to level the slurry with the top of the molds. See API RP 10B. 100072577 Revision H...
Use the cube mold removal tool to lower one or both molds into the Cement Curing Autoclave Chamber. Once the molds are in the chamber, avoid tipping the chamber, which may displace the covers. Figure 5-11 Lower mold into chamber – 1 Figure 5-12 Lower mold into chamber – 2 100072577 Revision H...
Install the lid onto the chamber. Fully tighten the lid into the chamber using the pressure vessel wrench and Vise Grip chain clamp. When the lid is fully seated, loosen it 1/8 to 1/4 turn. This helps to ensure the lid does not become jammed. 100072577 Revision H...
Water expelled as lid tightened 5.5.6.1 Attach the removable chamber handle by pulling the spring loaded pins away from the handle. Place the handle over the lid, and release the spring loaded pins to engage the holes in the lid. 100072577 Revision H...
Carefully lower the chamber into the heating jacket. Figure 5-18 Place chamber into autoclave, remove handle 5.6.3 Remove the carrying handle. 5.6.4 Connect the pressurizing manifold to the brass coupling on the chamber lid. Tighten the manifold coupling nut using hand force only. 100072577 Revision H...
Cement Curing Autoclave User Manual Figure 5-19 Attach valve and manifold 5.6.5 Connect the thermocouple cable to the thermocouple in the curing chamber lid. Figure 5-20 Connect thermocouple cable 100072577 Revision H...
Use the arrow keys to select the number of the program to be changed. Press. 5.8.2.3 For each program setting, use to change the displayed value. After releasing the button, the display will blink to show the controller has accepted the new value. 100072577 Revision H...
Hot water and/or steam may be violently expelled when the needle valve is opened. Keep all body parts clear of the valve outlet when it is being opened. Figure 5-22 Initial pressure (left), increased pressure from temperature rise (right) 100072577 Revision H...
Remove the molds from the chamber using the mold lifting tool. 5.10.6 Follow the procedures in “Curing at Pressures Above Atmospheric” in API RP 10B to remove the cured cement specimens from the molds, process, and test them. 100072577 Revision H...
5.11.4 Remove grease film with a solvent 5.11.5 Carefully dry the parts. 5.11.6 Inspect the lid O-ring for damage, and replace those which are damaged. 5.11.7 Clean and wipe all surfaces on the Cement Curing Autoclave. 100072577 Revision H...
The tables are grouped according to topic. The following troubleshooting tables are provided in this section: • General troubleshooting • Temperature troubleshooting • Pressure troubleshooting • Chamber lid troubleshooting 100072577 Revision H...
The system does not power Check/repair the power The power wiring is faulty. wiring. Refer to the wiring diagram. The power source is Check the power source. disconnected or turned off. Figure 7-1 Power fuse holders 100072577 Revision H...
The lid or chamber has a for damage or debris on the leak. surfaces, and clean or replace. The pressure gauge does not The pressure gauge is Test the pressure gauge, and zero. faulty. replace if found faulty. 100072577 Revision H...
The O-ring is damaged or Check and/or replace, and The lid is difficult to install. not properly lubricated. lubricate the lid O-ring. The O-ring seal area in Clean the chamber. chamber is dirty. 100072577 Revision H...
Cement Curing Autoclave User Manual 8 Accessories Hand Pressure Pump An optional-purchase hand-operated pump can pressurize the sample up to 3,000 PSI. Figure 8-1 101443558 PUMP ASSY HYD CEMENT CURING AUTOCLAVE, HAND OPERATED 100072577 Revision H...
For your protection, items being returned must be carefully packed to prevent ® damage in shipment and insured against possible damage or loss. FANN not responsible for damage resulting from careless or insufficient packing. Before returning items for any reason, authorization must be obtained from Fann Instrument Company.