PUMP ELECTRONIC CONTROLLER

SWITCH
MAINS

Technical Specifications

Input voltage: 200V-240V 50/60 Hz
Max input power: 1200 VA
Output voltage: 240 Vrms 3ph;
Max frequency: 64 Hz (factory setting)
Starting phase maximum power: 950 W
Normal Operation maximum power: 950 W
Maximum room temperature: 40 °C
Protection level: IP 20
CE mark:
- EN55011 cat "B"
- EN61000-3-2
- EN61000-3-3
- ESD EN61000-4-2
- EN61000-4-3
- EN6100-4-4
CSA mark:
-
EN61010-1, Installation category II,
pollution degree 2
-
EN60950
NOTE
Before connecting the IEC320 mains cable, install
the Retention Spring provided with the MS40+.
The Retention Spring has to be fastened to
IEC320 connector lateral screws.
NOTE
Use the Retention Spring to secure the mains ca-
ble into the IEC320 socket.
J1 SERIAL
P1 I/O 1
STATUS
LED
TANK
SENSOR
- EN61000-4-5
- RF EN61000-4-6
- EN61000-4-8
- EN61000-4-11
- EN61010-1

Input/Output Communications:

P1 – I/O
PIN N.
1
Interlock. Status
2
Speed setting (0-10V)
3
Oil level (relay contact)
4
Start (-)
5
Ground
6
Interlock. Status
7
Oil level (relay contact)
8
Start (+)
9
24V
Interlock: N.O. Relay contact – It is closed as
soon as the rotational frequency exceed the
threshold defined with window no.102 (plus
histerisys defined by window 105). Factory set
40Hz.
Speed setting: Electric motor speed setting –
linear setting between 1V=15Hz and 9V=
Win.120 setting:
-
High speed (60 Hz): Vin>9V
-
Low speed (40 Hz): Vin<1V
Oil level: N.O. Relay contact – It is closed as
soon as the oil level comes down under the
minimum level – The pump is not stopped but
the following starting is disabled.
Start/stop: If the inverter is managed by re-
mote port it manage the pump starting – if you
are operating in serial mode it enables the
starting. (pin 8 to pin 9 / pin 4 to pin 5).
Rotational speed setting by analog input (0-10V).
The pump operation is I/O signals dependent so
you must always plug-in the provided mating con-
nector to start the pump.
18/22
TECHNICAL INFORMATION
SIGNAL NAME
(N.O. relay contact)
(N.O. relay contact)
NOTE
87-901-008-01(A)
IN / OUT
out
in
out
in
out
out
out
in
out