2.4.1 Card installation

Perform the following steps, and refer to Figure 2-5 or Figure
2-6 to install the card assembly in the Model 7001/7002
mainframe.
WARNING
Turn off power to all instrumentation
(including the Model 7001/7002), and
disconnect all line cords. Make sure all
power is removed and any stored energy
in external circuitry is discharged.
1. Open the ejector arms at the back edge of the card.
2. Slide the card edges into the guide rails inside the main-
frame.
3. Carefully push the card all the way forward until the
ejector arms engage the mounting cups.
4. Push in on the card edge and ejector arms until the card
is properly seated.
5. Make sure the ejector arms are properly latched.
Figure 2-5
Card installation in Model 7001

2.4.2 Card removal

Follow the steps below to remove the multiplexer card from
the mainframe:
WARNING
Turn off power to all instrumentation
(including the Model 7001/7002), and
disconnect all line cords. Make sure all
power is removed and any stored energy
in external circuitry is discharged.
1. Pull out on the ejector arms until the card pulls free from
the internal connector.
2. Carefully slide the card out of the switching mainframe.
Ejector Arms (2)
Unlock card
Lock card