Advertisement

2-6.

CU17 CAPSULE ASSY

CAUTION
Take care so that you do not damage the CU17 capsule assy
with the wire terminal, etc. If you damage the CU17 capsule
assy, replace it with a new part.
capsule
holder (17)
Catch the wire onto the capsule holder (17).
2 lock ace screw (M1.4)
(L = 3.0 mm) (See Fig. B)
Note 3: When removing screw,
it is not reused.
Be sure to replace them
with new parts.
Note 1: When installing screws, follow the installing
procedure in the numerical order given.
tightening torque:
connector
side
1 terminal
4 CU17 capsule assy
capsule side
SIDE-002
CU17 capsule
board side
assy
1
CU17 capsule assy
Do not touch the portions within the dotted lines in the figure
below (including the wires connected to the capsule) with
your bare hands. If you do so, wipe your hands with alcohol.
Includes the capsule side portion and wire portion.
Make sure you do not touch these parts. If you do so, you
will not be able to use these parts. Make sure to replace
them with new parts.
SIDE-002 board side
3 lug
2 lock ace screw (M1.4)
(L = 3.0 mm) (See Fig. B)
Note 3: When removing screw, it is not reused.
Note 2: Align the lug with the groove
of the capsule holder (17) and
install.
2
connector
side
connector side
Be sure to replace them with new parts.
After installing the
two screws, refer to
"Wire setting" and
arrange the wire.
C-100
9