HP 8647A Operation And Service Manual page 133

Signal generator
Table of Contents

Advertisement

Required lkst
Equipment
w
HP 8903B (Option 051) Audio Analyzer
Distortion Accuracy: 1 1 dB
Residual distortion: -80 dB at 80 kHz BW
30 kHz low-pass filter
AC level accuracy: f4%
CCITT weighting filter
w
HP 8902A Measuring Receiver
FM accuracy: f 2 % of reading
1 1
digit
AM accuracy: f 2 % of reading f l digit
Frequency accuracy: 1 3 5 Hz at 1000 MHz (includes reference
Range: 250 kHz to 1000 MHz
Filters: 300 Hz high-pass; 15
kHz
low-pass
Detectors: Peak+
accuracy and counter accuracy)
Note
For the RF level accuracy performance test, the HP 8902A with a
HP 11722A sensor
is
required. (No other equipment substitutions are
allowed for this test.)
w
HP 8566B RF Spectrum Analyzer
Frequency range:
1 MHz
to 2000 MHz
Relative level accuracy: f 2 dB
w
HP 11722A Sensor Module
8-2 Performance
Tests

Advertisement

Table of Contents
loading

Table of Contents