7. Cable with gravity weight: Fasten cable with the pins, torque 7 Nm
(5.16 lbf ft)
Cable with centering weight: Fasten cable with the pins, torque
7 Nm (5.16 lbf ft) and fix the clamping part on the centering
weight.
8. Enter new probe length and then carry out a fresh adjustment
(see "Setup procedure, Carrying out min. adjustment - Carrying
out max. adjustment").
A
2
Fig. 27: Shortening the cable probe
A Gravity weight - cable ø 4 mm
B Gravity weight - cable ø 2 mm
C Centering weight - cable ø 2 mm
1
Threaded pins
2
Thread M8 for eye-bolt
3
Fixing screw - centering weight
9.7

Software update

The following components are required to update the sensor soft-
ware:
Sensor
Voltage supply
PC with PACTware
Current sensor software as file
You can find the actual sensor software as well as detailed infor-
mation of the procedure in the download area on our homepage:
www.siemens.com/sitranslg.
You can find information about the installation in the download file.
Caution:
Instruments with approvals can be bound to certain software versions.
Therefore make sure that the approval is still effective after a software
update is carried out.
You can find detailed information in the download area on our home-
page: www.siemens.com/sitranslg.