Operation; Remote Control (Option) - Siemens Mobilett Plus Installation Instructions Manual

Hide thumbs Also See for Mobilett Plus:
Table of Contents

Advertisement

Supplement to the Instructions for Use

Operation

Remote control (Option)

CAUTION
Take the remote control out of its remote holder. Step back from the X-ray unit
(up to 36+ feet / 10,9 m), and aim the remote control at the remote sensor on the front arm
of the X-ray machine.
The remote control is used in two-step. When pressing the exposure release switch once,
the preparation cycle is activated. The green "Ready" lamp goes out and does not light up
again until anode rotation has reached full speed (after approx. 2.5 s). This is indicated by
three short "beeps". The generator is now ready for exposure release.
NOTICE
To release the exposure, press the exposure release switch a second time, and hold it down
until the exposure is terminated. The amber exposure indicator lights during the exposure.
The indication lasts somewhat longer than the actual exposure, in order to make the oper-
ator aware of very short exposures. A longer beep indicates the end of the exposure.
After the exposure is completed, both the X-ray unit and remote control will automatically
return to stand-by mode.
Return the remote control back to its remote holder on the X-ray unit.
NOTICE
Siemens-Elema AB
Solna, Sweden
Make certain there are no other mobile units with a
TechSwitch remote system in line-of sight. This
includes any units behind a window or lead glass. If
there are other mobiles in line-of-sight, power down
those mobiles while operating the other.
It is possible to wait a maximum of 15 s in the prepara-
tion position before releasing the exposure, otherwise
the preparation will automatically be switched off.
If the remote control is not returned to its holder within
three minutes after use, the Remote finder feature will
activate causing the remote control to beep. The beeps
will continue indefinitely until the remote control is
located and put back in the remote holder.
SPR8-220.031.02
Rev. 02
Page 3 of 6
11.02
SPS-UD
4 - 3
4
4
Mobilett Plus/ E/ M/ HP
Service

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mobilett plus eMobilett plus mMobilett plus hp

Table of Contents