What materials do I need for proper cleaning?

Keysight N778-C Series Polarization Instruments User's Guide
Before you make any connection you must ensure that all cables and
connectors are clean. If they are dirty, use the appropriate cleaning
procedure.
When inserting the ferrule of a patchcord into a connector or an adapter,
make sure that the fiber end does not touch the outside of the mating
connector or adapter. Otherwise you will rub the fiber end against an
unsuitable surface, producing scratches and dirt deposits on the surface of
your fiber.
Some
Standard Cleaning Equipment
your instrument.
Standard Cleaning Equipment
Before you can start your cleaning procedure you need the following
standard equipment:
Dust and shutter caps
Isopropyl alcohol
Cotton swabs
Soft tissues
Compressed air
Dust and shutter caps
All of Keysight Technologies' lightwave instruments are delivered with
either laser shutter caps or dust caps on the lightwave adapter. All cables
come with covers to protect the cable ends from damage or
contamination.
We suggest these protected coverings should be kept on the equipment at
all times, except when your optical device is in use. Be careful when
reattaching dust caps after use. Do not press the bottom of the cap onto
the fiber too hard, as any dust in the cap can scratch or pollute your fiber
surface.
Isopropyl alcohol
This solvent is usually available from any local pharmaceutical supplier or
chemist's shop.
If possible avoid using denatured alcohol containing additives. Instead,
apply alcohol used for medical purposes.
Never try to drink this alcohol, as it may seriously damage your health.
Maintenance
on page 77 is necessary for cleaning