Keysight Technologies N778-C Series User Manual page 46

Polarization instruments
Table of Contents

Advertisement

3
Definition of Terms
Relative Power Uncertainty due to PDL
Relative Power Measurement Error due to Other Factors
Relative Output Power Uncertainty due to PDL
Relative Output Power Measurement Error due to Other Factors
Input Power Range
Maximum Safe Optical Input Power
WARNING
46
Two times the standard deviation of the power measured at the
polarimeter input when varying the polarization state (SOP).
Specifies the relative power measurement error when averaging out
the polarization dependent loss.
For each wavelength, the deviation of the measured power and the
input power is measured and the maximum relative deviation over all
wavelengths of the defined wavelength range is determined.
Two times the standard deviation of the measured output power error
when varying the polarization state (SOP).
The output power is measured at the (scrambler) output port (N7786C).
Specifies the relative power measurement error when averaging out the
polarization dependent loss.
For each wavelength, the deviation of the measured power and the output
power is measured and the maximum relative deviation over all
wavelengths of the defined wavelength range is determined.
The range of power levels for which the instrument displays a
measurement readout.
Higher power levels than defined by the maximum safe optical input power
may damage the instrument irreversibly. Note that proper operation is only
provided within the Input Power Range.
Higher power levels than defined by the maximum safe optical input
power may damage the instrument irreversibly.
Keysight N778-C Series Polarization Instruments User's Guide

Advertisement

Table of Contents
loading

This manual is also suitable for:

N7781cN7785cN7786cN7788c

Table of Contents