Electromagnetic Emissions; Electromagnetic Immunity - Braemar ER910 Manual

Arrhythmia event monitor
Table of Contents

Advertisement

ER910/920 and ER910/920 AF Event Monitor

Electromagnetic Emissions

Electromagnetic
Emissions test Compliance
environment - guidance
RF emissions
Group 1
The ER9xx uses RF energy only
CISPR 11
for its internal function. Therefore,
its RF emissions are very low and
are not likely to cause any
interference in nearby electronic
equipment.
RF emissions
Class B
The ER9xx is suitable for use in all
CISPR 11
establishments, including domestic
establishments and those directly
connected to the public low-
voltage power supply network that
supplies buildings used for
domestic purposes.

Electromagnetic Immunity

IEC 60601
Compliance
Immunity test
test level
level
Electrostatic
±6 kV contact
±6 kV contact
discharge (ESD)
±8 kV air
±8 kV air
IEC 61000-4-2
Power frequency
3 A/m
3 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
23
Electromagnetic
environment - guidance
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be at least
30%.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.