ER910/920 and ER910/920 AF Event Monitor
Additional equipment classification information as required in EN 60601-1
A. EQUIPMENT not suitable for use in the presence of a
FLAMMABLE ANAESTHETIC MIXTURE WITH AIR of WITH
OXYGEN OR NITROUS OXIDE
B. IPX0 Ordinary Equipment (enclosed equipment without protection
against ingress of water)
C. Internally Powered Equipment
D. Mode of Operation - Continuous Operation
21
ER910/920 and ER910/920 AF Event Monitor

Specifications

Functional
ER910
1 channel only
ER920
1 or 2 channel
Max Number of events
30
Sample rate
256 samples per second
User interface
LCD display and sound
Memory
Max event time
One channel
30 minutes
Two channel
30 minutes
Max total record time
One channel
30 minutes
Two channel
30 minutes
Type
Flash
Data retention
Non-volatile
Physical
Dimensions
3.5"x 2.125"x .65"
(89.9mm x 54.0mm x 15.7mm)
Weight with batteries
3.5 oz.
Enclosure
Molded plastic
Operating position
Any orientation
Electrical
Input impedance
2M min.
CMR ratio
60dB
AC signal range
+/- 3mV
DC signal range
+/- 300mV
Resolution
23uV (8bits)
Frequency response
.05Hz to 40Hz
Environmental
Operating temperature
0°C to +45°C
Non-operating temperature
-20°C to +65°C
Operating humidity
10% to 95% (non-condensing)
Non-Operating humidity
5% to 95%
Transtelephonic Transmission
Transmit carrier
1900Hz
Carrier deviation
100Hz/mV
Battery
Type
(2) AAA Alkaline IEC-LR3
Life
30 days min. during looping recording
Remove batteries during storage
Warranty
12 months from shipment