The CellarSafe is designed for easy installation using mains power supply. It is pre-calibrated and has been designed for one man installation. The alarm levels are factory set and the relays are ‘fail safe’ so in event of power failure the alarms will activate.
Page 5
CellarSafe is fitted with an AC plug for convenience, however if permanent installation is required, it should be wired to a dedicated AC spur fed from a circuit breaker, marked as the disconnecting device for the CellarSafe. The circuit breaker must comply with the relevant requirements of IEC 947-1and IEC 947-3. These procedures ensure that the installation complies with the requirements EN 61010-1 (Low Voltage Directive).
Page 7
3.0% Note: CO2 alarms are Subject to a 25 second time delay to prevent false alarms when exchanging gas cylinders. Oxygen alarms are instantaneous. 115v ac option available Issue 5...
To test the alarms, blow into a plastic bag and hold the open end of the inflated bag tightly over the sensor(s) until the alarms activate. The CellarSafe display will show increased levels of CO and decreased levels of oxygen (on units with an oxygen sensor fitted), and the alarms should...