Hold Percentage
The Hold Percentage provides just enough power to the heating pads to compensate for heat loss, clamping
the temperature at the set point. The temperature hold heating percentage is used when the temperature is
within +/- 1 degree of the Set Point temperature.
The actual Hold Percentage is dynamically adjusted according to the measured temperature. When the
temperature settles below the Set Point, the adjusted percentage is incremented. When the temperatrue
settles above the Set Point, the adjusted percentage is decremented. These small changes in power will
slowly nudge the temperature towards the Set Point, while minimizing over/under shooting.
The Hold Percentage setting remains unaffected by the adjusted setting. The adjusted setting remains in
affect until the heater control box is powered off. A good starting value is 10 Percent.
6.2
Inactivity Shut Off Timer
When enabled, the heater will automatically exit Active Mode (stop heater) after set number of hours of
inactivity from the keypad or remote communications. The default is disabled (0 hours setting).
After selecting, use the arrow buttons to set the inactivity shut off time in hours. Set to 0 to disable timer.
The set time will be stored in non-volatile memory.
nn
During keyapd or remote inactivity while the heater is active, the display will periodically display the
inactivity time remaining until the heater exits Active Mode and stops heating the pad. The time will be
displayed in Hours and Minutes, or Minutes and Seconds if less then 1 hour remains.
nn nn
6.3

Lock Out Mode

When enabled, Lock Out mode will lock out the keypad until the user defined lock out code is entered.
If any button is pressed while locked out, the display will show:
Pressing either arrow button will enter Lock Out code entry mode:
return to the previous display.
Use the arrow buttons to set the blinking digit. Press the Setup or Enter buttons to move to the next digit.
If the code entered is wrong, the display will show:
If the code is correct, the keypad will be unlocked, but the Lock Out setting will be unchanged. The next
time power is turned on, the keypad will be locked out. Use Setup mode to disable Lock Out mode.
RS-232 communications is not affected by Lock Out Mode.
The Lock Out code can be remotely set or queried with the Lock Out command:
LOC [ < 0 | 1 [<nnnn>] >
Publication #1200-01
www.SyringeHeater.com
6
Model HEATER-KIT-1
nnnn
. Any other button will
. Press any button to clear.
12/25/20