User Interface; Setup Settings - NEWERA Pump Systems Syringe HEATER-KIT-1LG Complete User's Manual

Table of Contents

Advertisement

New Era Pump Systems Inc.

User Interface

4.
The controller will display, as the default display, the current
temperature of the primary heating pad, in the currently selected units,
C or F.
LEDs indicate the status of the heater.
Temperature Set Point is actively being
ACTIVE
maintained
Heating pads are being powered. Flashing
OUT
indicates on/off power duty cycle to slow down
heating when near the Set Point
Heating pad temperature is greater than 20
ALARM
degrees C above the Set Point. Other system
alarms.
SETUP
ENTER
STOP
┌ ACTIVE ┐
ENTER
SETUP
STOP

Setup Settings

5.
Press the SETUP button
display the next menu entry. All settings are saved in non-volatile memory. Pressing an Up or Down
arrow button will immediately select Temperature Clamping Set Point and will increment or decrement the
setting.
Press the ENTER button
Configuration. A value being changed will be automatically entered after a few seconds time out period.
The display will blink to indicate the value has been entered and saved in non-volatile memory.
Publication #1200-01
www.SyringeHeater.com
Keypad
Increases or decreases settings
By default, will change the temperature Set Point
Settings with decimal points
Press both buttons simulaneously to shift the decimal point.
Displays the setup menu. Each press will display the next menu selection.
When displaying the setup menu or changing a value, will select the entry or
save the new value. The display will blink when the new value is saved in
non-volatile memory.
Otherwise, will stop the heater. The Active LED will turn off. Temperature
will no longer be maintained.
Simultaneously press the SETUP and ENTER/STOP buttons. The Active LED
will turn on and the heatings pads will start heating until the temperature
reaches the Set Point.
SETUP
to display the Setup Menu. Each press of the SETUP button will
ENTER
STOP
to select a menu entry or accept and enter a setting, exiting Setup
SETUP
ALARM
Up / Down Arrow Buttons
SETUP Button
ENTER / STOP Button
Active Mode Buttons
3
Model HEATER-KIT-1
ACTIVE
ENTER
STOP
ACTIVE
OUT
12/25/20

Advertisement

Table of Contents
loading

This manual is also suitable for:

Syringe heater-kit-5sp

Table of Contents