1. Safety Instructions 1.1 Correct Usage • The EF200 measuring system is used to measure the flow of saturated steam, superheated steam, air and water. Do not use to measure the flow of toxic, flammable or otherwise hazardous fluids. Use this system only as intended.
• Observe all local regulations governing the opening and repair of electrical devices. 1.5 Repairs, Dangerous Chemicals Warning! The following procedures must be carried out before an EF200 is sent to TLV for repair: Warning! NOTE: References to use with hazardous fluids are for customers having special permission and a signed contract with TLV for hazardous use.
-200 to +400 °C and at nominal pressures of up to 4.96 MPaG (49.6 barg). EF200 can measure the volumetric flow rate in operation and can be programmed to supply the flow rate in mass, energy or corrected volume units via temperature measurements by the internal temperature sensor.
IP66/67, Type 4X enclosure Temperature Ranges • The maximum allowable ambient and process temperatures must be observed (see 6.1.6). • Ensure both pipeline heat insulation and mounting position conditions are met (see 3.3). 172-65639M-01 (EF200) 13 Apr 2018...
A vortex flowmeter requires a fully developed flow profile as a prerequisite for measuring volume accurately. The following points must therefore be noted when mounting the EF200 in the pipeline. Pipe Inner Diameter When ordering, ensure that the nominal diameter and pipe schedule...
3.3.3 Installation Orientation The EF200 can be mounted in any direction in the piping. An arrow on the meter body shows the direction of flow. For measuring liquids in vertical pipes, the meter should be installed with an upwards flow...
Caution! Removing the transmitter from the housing support is to be carried out by qualified TLV appointed service personnel only! Caution! 3.3.7 Ensuring accurate measurements In some cases where steam is mixed with condensate, it may not be possible to obtain accurate flow rate measurements.
When installing a bypass line, use upstream and downstream valves of a type that does not disturb the flow profile, and secure sufficient length of straight pipe. Bypass Valve Direction of Flow Ball Valve Ball Valve EF200 172-65639M-01 (EF200) 13 Apr 2018...
Gaskets that protrude into the flow affect vortex formation behind the bluff body and lead to inaccurate measurement. Therefore, the gaskets delivered by TLV come with a slightly larger inner diameter than the measuring pipe.
(optional) Unit: mm Caution! Take care not to overtighten the nuts when installing the remote transmitter on piping. Caution! The transmitter housing is constructed from die-cast aluminum, therefore may be damaged. (Recommended torque: 5 N⋅m) 172-65639M-01 (EF200) 13 Apr 2018...
3.6 Transmitter Housing / Display (Mounting/Rotating) Rotating the Transmitter Housing The transmitter housing of EF200 can be rotated on the housing support up to 350° clockwise or counterclockwise to reposition the optional local display for easy reading. This is carried out as follows (see Fig. 11): 1.
If installation outdoors in an uncovered location in unavoidable, installing the optional sunshade is recommended. (This is not required when installing compact version with the transmitter oriented downwards.) Figure 13 Installing the optional sunshade 172-65639M-01 (EF200) 13 Apr 2018...
Procedure for connecting the transmitter To remove a cable from the terminal, use a flat-blade screwdriver to push the slot between the two terminal holes while simultaneously pulling the cable end out of the terminal. 172-65639M-01 (EF200) 13 Apr 2018...
4.2 Wiring Diagrams EF200 Only, Connecting Power Supply Power Supply – +24V Figure 15 Power Supply DC24V Frequency (Pulse) Output for an Electronic Counter or PLC Electric Counter or PLC (Open collector input) Power Supply – +24V Figure 16 Pulse output to electronic...
4.3 Connecting to TLV EC351 Flow Computer and Parameter Settings Figure 18 Connecting to TLV flow NOTE: - Assign temperature to the analog output computer EC351 - Shielded cable recommended for analog output EF200 Pulse Output Settings Submenu Parameter Selection...
Terminal 3 = yellow cable Terminal 4 = green cable 5. Correct the cable shield via the cable strain relief. 6. Screw the housing cover back onto the transmitter housing, then reposition the securing clamp and tighten. 172-65639M-01 (EF200) 13 Apr 2018...
Page 22
For non-standard specifications, there may be cases in which the signal cable has no plug and requires terminal connection identical to the sensor housing. Caution! Consult TLV for cable connection methods. Connection Cable Specifications 4 × 2 × 0.34 mm...
5. Operation The EF200 has a number of functions that the user can individually set according to process conditions. The display consists of max. of 4 lines; this is where measured values and/or status variables are displayed. You can change the assignment of the display lines to different variables to suit your needs and preferences.
The direct access code for the parameter you are navigating to (e.g. 0022-1) – If a diagnostic event is present, the diagnostic behavior and status signal – • Of the wizard: If a diagnostic event is present, the diagnostic behavior and status signal – 172-65639M-01 (EF200) 13 Apr 2018...
Page 25
Expert Appears: • In the menu next to the "Expert" section • At the left in the navigation path in the "Expert" menu NOTE: This menu is for use by TLV technical service personnel only. Submenus, wizards, parameters Symbol Meaning...
Selection of letters from a to z Selection of special characters Confirms selection Switches to the selection of the correction tools Exits the input without applying the changes Clears all entered characters Correction keys under 172-65639M-01 (EF200) 13 Apr 2018...
Reduces the contrast (brighter setting). Plus/Enter key combination (press and hold down the keys simultaneously) Increases the contrast (darker setting). Minus/Plus/Enter key combination (press the keys simultaneously) – For operational display Enables or disables the keypad lock. 172-65639M-01 (EF200) 13 Apr 2018...
• Operation • Setup • Diagnostics • Expert Example: Setting the number of displayed measured values to “2 values” The mark indicates the current setting. Press for 2 seconds 172-65639M-01 (EF200) 13 Apr 2018...
Keylock on appears. Switching off the keypad lock The keypad lock is switched on. Press the keys simultaneously. – The message Keylock off appears on the display: The keypad lock is switched off. 172-65639M-01 (EF200) 13 Apr 2018...
100 Hz • Switch output: Binary, conductive or non-conductive Can be set to status, limit values for all measured variables or diagnostic behavior Signal on Alarm Output behavior for alarm can be configured for current/pulse/frequency/switch output 172-65639M-01 (EF200) 13 Apr 2018...
DC sparkover voltage : 400 to 700 V Trip surge voltage : <800 V Capacitance at 1 MHz : <1.5 pF Nominal discharge current : 10 kA (8/20µs) : -40 to +85 °C Temperature range 172-65639M-01 (EF200) 13 Apr 2018...
• Remote Version – Sensor: -40 to +80 °C Display can be read between -20 °C and +60 °C Caution! When mounting outside, TLV recommends to protect the device from direct sunlight with a sunshade (optional part), Caution! especially in warmer climates with high ambient temperatures.
ASME class 150, 300 [bar] -200 -100 0 100 200 300 400 [°C] Limiting Flow See flow rate data in section 11.1 Pressure Loss TLV will calculate and provide pressure loss data on request 6.1.7 Mechanical Construction Dimensions See 6.2 to 6.5 Weight See 6.2 to 6.5...
Remote Operation Can be operated using the following tools. • HART communication 6.2 Remote Transmitter Dimensions Electrical Cable Connection LED Display Connecting Cable Port Figure 23 Dimensions of transmitter, remote Weight: 2.4 kg Unit: mm version 172-65639M-01 (EF200) 13 Apr 2018...
Totalizer 1 to 3 NOTE: • TLV will preconfigure the parameters necessary for measurement after determining the customer’s specification. Note! • See 4.3 for more information when using EF200 in combination with TLV EC351 Flow Computer. 172-65639M-01 (EF200) 13 Apr 2018...
Simulates various process variables, device alarm mode and Simulation 8.2.25 signal output without a real flow situation For use by TLV technical Expert 8.2.26 personnel. TLV does not disclose access codes to users. 172-65639M-01 (EF200) 13 Apr 2018...
If an incorrect access is entered, the user obtains the access rights of the “Operator” role. NOTE: The “Expert” status is assigned only to TLV technical personnel. TLV does not disclose access codes to users. 172-65639M-01 (EF200) 13 Apr 2018...
Specify start value for totalizer. (Signed floating-point number) Preset value 1 to 3 Factory setting: 0 m Reset all totalizers to 0 and start. Reset all totalizers • Cancel • Reset + totalize Factory setting: Cancel 172-65639M-01 (EF200) 13 Apr 2018...
NOTE: - In the Select gas type parameter, the User-specific gas option is selected, or - In the Select liquid type parameter, the User-specific liquid option is selected. • Heat • Calorific value Factory setting: Heat 172-65639M-01 (EF200) 13 Apr 2018...
• Min. • Max. • Last valid value • Actual value • Defined value Factory setting: Max. Enter current output value in alarm condition. Failure current Setting range: 3.59 to 22.5 mA Factory setting: 22.5 mA 172-65639M-01 (EF200) 13 Apr 2018...
• No pulses Factory setting: No pulses NOTE: This function is available when the operating mode is Invert output signal set to “pulse output”. Invert the output signal. • No • Yes Factory setting: No 172-65639M-01 (EF200) 13 Apr 2018...
Page 49
Setting range: 0.0 to 1250.0 Hz Factory setting: 0.0 Hz NOTE: This function is available when the operating mode is Invert output signal set to “frequency output”. Invert the output signal. • No • Yes Factory setting: No 172-65639M-01 (EF200) 13 Apr 2018...
Page 50
Factory setting: Depends on customer specifications. NOTE: This function is available when the operating mode is Switch-off value set to “switch output”. Enter measured value for the switch-off point. (Signed floating-point number) Factory setting: Depends on customer specifications. 172-65639M-01 (EF200) 13 Apr 2018...
Enter on value for low flow cut off. (Positive floating-point On value low flow number) cutoff Factory setting: 0 Enter off value for low flow cut off. Off value low flow cutoff Setting range: 0 to 100.0% Factory setting: 50% 172-65639M-01 (EF200) 13 Apr 2018...
Factory setting: Depends on customer specifications. Corrected volume Select corrected volume unit. (The selected unit is taken from unit corrected volume flow unit parameter.) • Nl, Nm , Sm , Sft , UserCrVol Factory setting: Depends on customer specifications. 172-65639M-01 (EF200) 13 Apr 2018...
Page 54
• Pa s, cP, P Factory setting: Pa s Select length unit for nominal diameter. Length unit The selected unit is taken from inlet run and mating pipe diameter. • mm, m, in, ft Factory setting:mm 172-65639M-01 (EF200) 13 Apr 2018...
Enter gross calorific value to calculate the energy flow. Calorific value (Positive floating point number) Factory setting: 50 000 kJ/kg Enter real gas constant Z for gas under operation conditions. Z-factor Setting range: 0.1 to 2.0 Factory setting: 1 172-65639M-01 (EF200) 13 Apr 2018...
• 2nd temperature delta heat Factory setting: Off Enter atmospheric pressure value to be used for pressure Atmospheric correction. pressure (The unit is taken from the pressure unit parameter.) Setting range: 0 to 250 bar Factory setting: 1.01325 bar 172-65639M-01 (EF200) 13 Apr 2018...
(Example: Cannot be used in conjunction with the “Steam Dryness Fraction Calculator” option). Factory setting: Off Define length of the straight upstream piping. Inlet run Setting range: 0 to 20 m Factory setting: 0 m 172-65639M-01 (EF200) 13 Apr 2018...
Set display reaction time to fluctuations in the measured value. Display damping Setting range: 0.0 to 999.9 seconds Factory setting: 5 seconds Select header contents on local display. Header • Device tag • Free text Factory setting: Device tag 172-65639M-01 (EF200) 13 Apr 2018...
Comparison between present device data and display backup. Comparison result • Settings identical • Settings not identical • No backup available • Backup settings corrupt • Check not done • Dataset incompatible Factory setting: Check not done 172-65639M-01 (EF200) 13 Apr 2018...
NOTE: Using the Device reset parameter it is possible to reset the entire device configuration or some of the configuration to a defined state. • Cancel • To factory defaults • To delivery settings • Restart device Factory setting: Cancel 172-65639M-01 (EF200) 13 Apr 2018...
(e.g. @, %, /) Factory setting: EF200 Serial number Displays the serial number of the measuring device. Max. 11-digit character string comprising letters and numbers. Firmware version Displays the device firmware version installed. Display format: xx.yy.zz 172-65639M-01 (EF200) 13 Apr 2018...
In the assigned Process variable parameter of Totalizer 1 to 3, the current totalizer counter value and the current totalizer overflow can be displayed. Totalizer value: Current totalizer counter value Totalizer overflow: Current totalizer overflow 172-65639M-01 (EF200) 13 Apr 2018...
Switch simulation of the frequency output on and off. Frequency • Off simulation • On Factory setting: Off Enter the frequency value for simulation. Frequency value Setting range: 0.0 to 1 250.0 Hz Factory setting: 0.0 Hz 172-65639M-01 (EF200) 13 Apr 2018...
(depends on the selected category) Factory setting: Off 8.2.26 Expert This menu is for use by TLV technical service personnel and does not require operation during standard flow measurement. In the event that operation of this menu is required, TLV will explain the operation procedure.
• Outside its technical specification limits (e.g. outside the process temperature range) • Outside of the configuration carried out by the user (e.g. maximum flow in parameter 20 mA value) Maintenance required Maintenance is required. The measured value remains valid. 172-65639M-01 (EF200) 13 Apr 2018...
Page 68
2. Select the desired diagnostic event with and press – The message for the remedial measures for the selected diagnostic event opens. 3. Press simultaneously. – The message for the remedial measures closes. 172-65639M-01 (EF200) 13 Apr 2018...
One year following product delivery. 2. Warranty Coverage TLV CO., LTD. warrants this product to the original purchaser to be free from defective materials and workmanship. Under this warranty, the product will be repaired or replaced at our option, without charge for parts or labor.