Page 1
Technical Note: Installing the Alarm Box Add-On on the Cirris 1000HX Cable Analyzer Version 3.0 September, 1998...
Page 2
Technical Note: Installing the Alarm Box Add-On on the Cirris 1000HX Cable Analyzer Version 3.0 September, 1998 Copyright 1998 by Cirris Systems Corporation All Rights Reserved Cirris Systems Corporation 1991 Parkway Boulevard Salt Lake City, Utah 84119-2026 United States of America...
Page 3
How to contact us If you find you don’t have one of the items you’ll need, or you have other questions, please feel free to telephone us at 1-801-973-4600 or 1-800-441-9910. Our fax telephone number is 801-973-4609. 1000HX Alarm Box Add-On Tech. Note page 1...
In this section, we’ll give you a step-by-step procedure for installing the Alarm Box here? Add-On accessory on your Cirris 1000HX. To make things easier, we’ll provide digital photographs taken at the workbench, as one of our factory technicians installed an alarm box.
Page 6
Disconnect the ribbon cable from the scanner assembly, set the scanner assembly aside. Place the Alarm Box next to the 1000HX. Note: When you latch the two boxes together shortly, the pins on the side of the Alarm Box will fit into the holes in the side of the analyzer itself.
Page 7
Turn the Processor Assembly over gently. You’ll see the ribbon cable still connected to it. The wires running to the system speaker are also still connected. Ribbon Cable (still connected) Speaker Wires (still connected) 1000HX Alarm Box Add-On Tech. Note page 5...
Page 8
In this view, the Processor Assembly and the Scanner Assembly have both been laid aside lying on their tops. The analog assembly with its metal frame has been lifted out of the chassis, and set on the bench. 1000HX Alarm Box Add-On Tech. Note page 6...
Page 9
Replace the top assembly on the alarm box. Note that the long ribbon cable should snake across the entire bottom of the analyzer’s chassis box. Long Ribbon Cable Speaker Wires (disconnected) 1000HX Alarm Box Add-On Tech. Note page 7...
Page 10
Center Cover Plate Scanner Assembly Scanner Cover Plate Wide Ribbon Cable Analog Assembly Speaker Wires Long ribbon cable Analog-to-Processor Cable (Wide Ribbon Cable) Speaker Wires and Long Ribbon Cable 1000HX Alarm Box Add-On Tech. Note page 8...
Page 11
Analog-to-Processor ribbon cable, the narrower long ribbon cable, and the speaker wires should all be connected to the processor assembly. If they’re not all connected as shown in this photo, connect them. Narrow Ribbon Cable Speaker Wires Analog-to-Processor wide ribbon cable 1000HX Alarm Box Add-On Tech. Note page 9...
Page 12
Be sure to lower the assembly at an angle as shown! This will avoid snapping the pogo pin that grounds the processor to the chassis. 14. Replace the center cover plate as shown here. 1000HX Alarm Box Add-On Tech. Note page 10...
Page 13
Reconnect the Processor-to- Scanner ribbon cable... 16. Replace the scanner cover plate. The installation of the Alarm Box is now complete! Replace the Scanner Cover Plate 1000HX Alarm Box Add-On Tech. Note page 11...