Alarm / Condition - BD Alaris GW 800 Series Quick Reference Manual

Volumetric pump
Hide thumbs Also See for Alaris GW 800 Series:
Table of Contents

Advertisement

An alarm indicates that the infusion has stopped.
Check the display for an alarm symbol.
Press
to silence the sound. (Exceptions are
c
silenced)
When the cause has been rectified, press
All alarms will stop the infusion except for
where the infusion status will remain the same state as prior to
the alarm.

Alarm / Condition

J
Air-in-line
Upstream occlusion
K
Internal battery depleted.
L
Door open
M
System fault
N
Flow error
Flow sensor attached to secondary
infusion set.
O
Flow sensor connection error.
P
Downstream occlusion
Q
Incorrect infusion set, infusion set
incorrectly loaded or infusion set worn.
and
M
K
key to resume infusion.
d
F
Action
• Disconnect the infusion set from the
IV access device and use the prime
function.
• Remove occlusion/air and restart the
infusion.
• Connect pump to AC power.
• Ensure pump is switched on to charge
internal battery.
• Close door and restart infusion.
• Switch pump off.
• Remove pump from service and have
the pump inspected by a qualified
service engineer.
• Close in-line clamp.
• Ensure tubing is loaded correctly.
• Ensure ample fluid is in container.
• Clear blockage. Close pump door.
• Restart infusion.
• Ensure flow sensor is attached to
primary infusion set.
• Connect flow sensor to pump and
restart infusion. Ensure fluid in drip
chamber is not above fill line.
• Remove pressure in the infusion set to
prevent a post occlusion bolus to the
patient.
• Remove the cause of the blockage.
• Restart the infusion.
• Use infusion set as specified in DFU.
Ensure infusion set is loaded correctly
into pump.
0000CF04392 Iss. 1 9/12