Thermometers are used for measuring body temperature by oral, axillar and rectal method.
Although the generally accepted "normal"temperature reading is 36.6˚C, temperature readings can vary from 35.2˚C to 36.8˚C and still may be considered "normal". Variations in temperature can be attributed to activities such as exercise, smoking, eating and drinking. Your temperature is lower in the morning than in the afternoon. Other variations may be due to the location of the temperature reading: Axillary (armpit measuring), Oral (mouth measuring), Rectal (rectum measuring). While the axillary temperature follows the guidelines previously described, an oral temperature is generally 0.5˚C higher. Rectal temperature will be 1.0˚C higher.
Press ON/OFF button to activate. The unit would beep and a display would show all the segments for around 2 seconds.
WT-03 | WT-04 | WT-06 | |
Axillary | 2-3 min | 50-60 sec | 20 sec |
Oral | 1 min | 40-50 sec | 10-15 sec |
Rectal | 1 min | 30-40 sec | 6-10 sec |
Taking an axillary temperature may take longer time because of the features described above. When the temperature rises slightly the unit beeps.
Attention! When a unit beeps it doesn't mean that measuring is completed. It only means that your temperature is rising slightly. Keeping the unit for 3060 sec after the beep is recommended.
Beeps definition:
WT-03 | WT-04 | WT-06 | |
Short | t ≥ 37,5˚C | t ≥ 37,8˚C | t ≥ 37,8˚C |
Long | t <37,5˚C | t <37,8˚C | t <37,8˚C |
In order to prolong the battery life, turn OFF the unit after use pressing ON/OFF button. The unit will automatically turned OFF approximately after 10 minutes.
Note: According to hygiene standards do not use for rectal measuring the same unit as used for axillary or oral measuring. Avoid having hot or cold drinks, exercising, smoking or taking a shower directly before measuring.
Note: The unit may be damaged if
Contraindications: are not revealed.
The battery in your digital thermometer needs to be replaced when "" or "
" appears in the lower right corner of the display.
To replace the battery:
Note: Please keep the battery away from children. Do not store the battery at a high temperature.
WT-03 | WT-04 | WT-06 | |
Measurement range | 32,0 ˚C - 44,0 ˚C | 32,0 ˚C - 42,9 ˚C | 32,0 ˚C - 42,9 ˚C |
Display resolution | 0,1 ˚C | 0,1 ˚C | 0,1 ˚C |
Accuracy | Accuracy ±0,1˚C for measurement range 35,5˚C - 42,0˚C Accuracy ±0,2˚C for measurement range lower than 35,5˚C and higher than 42,0˚C | Accuracy ±0,1˚C for measurement range 35,5˚C - 42,0˚C Accuracy ±0,2˚C for measurement range lower than 35,5˚C and higher than 42,0˚C | Accuracy ± 0,1˚C for measurement range 35,5˚C - 42,0˚C Accuracy ± 0,2˚C for measurement range lower than 35,5˚C and higher than 42,0˚C |
Auto power OFF | 10 min ±2 min | 8 min 40 sec ±2 min | 10 min ±2 min |
Battery | LR41 / SR41 | LR41 / SR41 | LR41 / SR41 |
Life of battery | Min 1 year | Min 1 year | Min 1 year |
Unite size | 128,1 х 18,5 х 10,5 mm | 133 х 21 х 11 mm | 148 х 22 х 20 mm |
Weight (with battery, without case) | ~ 10 g | ~ 10 g | ~ 14 g |
Operating condition | Temperature: 10°C ~ 40°C; relative humidity: 30% ~ 85% | ||
Storage condition | Temperature: -10°C ~ 60°C; relative humidity: 25% ~ 90% |
The unit must be utilized in accordance with current standards separately from domestic wastes. For utilizing it is necessary to contact special organizations licensed to make utilization.
This unit meets EC requirements for medical equipment 93/42/ЕЕС. Declaration of conformity.
Note: You may check the unit accuracy yourself. Pour a vessel (300-500 ml) with warm water from 36 to 41°С. Put two or more thermometers including one you want to check and a reference one.
Probe tips must be at the same level and 2-3 cm in water. After 1-2 min compare the figures. Their diference must be not more or less than 0,1°C.
Warranty period is 2 years from the date of purchase. This warranty doesn't cover any damages caused by improper using, and also battery, protective cover and packaging. When a manufacturing defect is revealed during the warranty period a faulty unit would be repaired or, if repairing is impossible, replaced with another one.
Manufacturing date is on the unit.
LOT: month number, last 2 figures of the year.
The manufacturer may change units partially or completely if necessary, without prior notice.
FOLLOW INSTRUCTIONS BEFORE USE
HOUSING INGRESS PROTECTION RATE: IP22
(Protected from the penetration of solid bodies with dimension greater than 12.5 mm. Protected from the penetration of vertically falling water drops)
MANUFACTURER'S NAME
CE MARK
TYPE BF EQUIPMENT
ARTICLE NUMBER
LOT NUMBER
OPERATING CONDITION, TEMPERATURE
STORAGE CONDITION, TEMPERATURE
B.Well Swiss AG
Bahnhofstr. 24, 9443 Widnau, Switzerland
www.bwell-swiss.ch
Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.
Download SteriPEN WT-03 BASE, WT-04 STANDARD, WT-06 FLEX Manual
Advertisement
Need help?
Do you have a question about the WT-03 BASE and is the answer not in the manual?
Questions and answers