MED-3000 advantages include:
The device is intended for the intermittent measurement and monitoring of human body temperature from forehead measurement at home, clinics and hospital.
A control measurement using a conventional thermometer is recommended in the following cases:
When using this product, please be sure to follow all the notes listed below. Any action against these notices may cause injury or affect the accuracy.
and
As with any thermometer, proper technique is crucial to obtaining accurate temperature readings.
Please read this manual thoroughly and carefully before using.
Contraindications: are not revealed.
The temperature in the forehead and temple area differs from the internal temperature, which is taken orally or rectally.
Vasoconstriction, an effect which constricts the blood vessels and cools the skin, can occur during the early stages of a fever. In this case, the temperature measured by the MED-3000 Infrared thermometer may be unusually low. If the measurement therefore does not match the patient's own perception or is unusually low, repeat the measurement every 15 minutes. As a reference, you can also measure the internal body temperature using a conventional oral or rectal thermometer. Body temperature can vary from one individual/ person to next. It also varies by location on the body and time of day. Below show the statistical normal ranges from different sites. Please keep in mind that temperatures measured from different sites, even at the same time, should not be directly compared. Fever indicates that the body temperature is higher than normal. This symptom may be caused by infection, overdressing or immunization. Some people may not experience fever even when they are ill. These include, but are not limited to, infants younger than 3 months old, persons with compromised immune systems, persons taking antibiotics, steroids or antipyretics (aspirin, ibuprofen, acetaminophen), or persons with certain chronic illnesses. Please consult your physician when you feel ill even if you do not have fever.
Normal Temperature Range of various body sites
Oral | 36,4 - 37,6°C |
Rectal/ Ear | 36,7 - 38,2°C |
Axillary | 35,8 - 37,3°C |
Low battery warning:
The thermometer comes with two 1.5 V AAA alkaline batteries. The thermometer will display " " to alert you when its power is getting low. Please follow the steps below to replace new batteries immediately.
Replacing the Battery:
NOTE:
This thermometer can display results in either degrees Celsius (°C) or degrees Fahrenheit (°F).
This thermometer provides Tracking Light (LED Light) to help users placing the thermometer at the correct position.
Following modes can be selected:
Choose the mode you need using a slider on the side of the device.
Silent mode
Beeps may be switched off:
Tips for measuring human temperature
Bear in mind that the thermometer needs to have been in the room in which the measurement is taken for at least 30 minutes before use.
NOTE:
Measuring Human Temperature
Taking a Measurement
Taking a Measurement
Memory Recall
You can recall up to 25 measurements currently stored in memory to share with your physician or trained healthcare professional.
Memory Deletion
Automatically on the 26th measurement: when the 25 memories have been used up, any new measurement will be recorded and the oldest memory deleted without you having to do anything.
NOTE:
All the readings will be cleared no matter record in Body Mode or Object Mode.
For the device disinfection, 75% alcohol (available in the pharmacy) can be used.
Measurement Sensor:
Please wipe with a clean and dry cotton swab to clean the Sensor on the inside. Do not use water, the Sensor is not waterproof.
Thermometer:
Use a soft and dry cloth to clean the thermometer body. Do not use abrasive cleaners. The body of the thermometer is not water-resistant. Never put the thermometer under a running tap or submerge it into water.
Store the thermometer in a cool and dry location. Free from dust and away from direct sunlight. Do not expose the thermometer to extreme temperatures, humidity, or shock.
FAULT OR FAULT SYMBOL | FAULT DESCRIPTION | CORRECTIIVE MEASURE |
No display On the LCD pane. | The battery has run out. Incorrect battery polarity. | Replace the battery. Please note: The (+) side of the battery must face Positive (+) indication on the device. |
Measurement not possible (or an abnormal value is displayed). | The thermometer is not ready. | Wait until the °C symbol is displayed. |
An abnormal temperature value is displayed. | The Sensor is dirty or damaged. | Clean the Sensor or get it repaired. |
The thermometer was removed too early (before the beep sound). | Wait until you hear the beep before removing the thermometer from the forehead. | |
symbol is displayed. | The temperature measured is outside the measuring range. Body Mode: Object Mode: | Check that the probe tip is clean and that the thermometer is properly placed on the forehead. |
Symbol is displayed. | The battery has run out. | Replace the battery. |
Symbol is displayed. | The ambient temperature is outside the operating temperature range or is changing too fast. | To ensure accurate measurement, let the thermometer rest at operating temperature for 30 minutes prior to use. |
Device Standards:
Device Corresponds to the requirements of the standard for infrared thermometers
IEC 60601-1:2012
IEC 60601-1-2:2014 ISO 80601-2-56:2009
Classification:
Anti-electric Shock Type: Internally powered equipment
Applied part: Type BF
Mode of operation: Continuous Operation
EMC: type B class I
Electromagnetic Compatibility:
Device fulfi lls the stipulations of the standard IEC 60601-1-2
The stipulations of EU-Directive 93/42/EEC for Medical Devices
Class Ila have been fulfi lled.
Measurement range: | |
Human Body: | 32,0 ~ 42,9°C (89,6°F ~ 109,3°F) |
Object: | 0,0°C ~ 100°C (32,0°F ~ 211,8°F) |
Accuracy: | |
For Body mode: | ±0,2°C (±0,4°F): 36,0°C ~ 39,0°C (96,8°F ~ 102,2°F) ±0,3°C (±0,5°F): out of the range |
For Object mode: | ±1°C (±2°F): 0°C ~ 60,0°C (32,0°F ~ 140,0°F) ±4°C (±7,2°F): out of the range |
Display resolution: | 0,1°C or 0,1°F |
Temperature unit: | Default °C or Default °F |
Operation Distance: | 3 cm |
Traffic lights backlight (Body mode), Indicator light range: | Green: <37,5°C (<99,5°F); Yellow: ≥37,5°C - ≤37,9°C (≥99,5°F - ≤100,2°F); Red: >37,9°C (>100,2°F) |
Beeps defi nition: | Power on and ready for working: a short beep Measurement finished: 1 long beep ≤37,9°C (100,2°F) 6 short beeps >37,9°C (100,2°F) |
Memory: | 25 mesurements |
Auto shut-off: | about 10 seconds after no using |
Batteries: | 2 x 1.5V AAA size alkaline batteries |
Dimensions (L×W×H): | 142,5 mm × 38 mm × 40 mm |
Weight: | 76g (Net: 54g) |
Operating environment: | Temperature: +5°C ~ +40°C (41°F ~ 104°F); Humidity: 15% ~ 93% RH (non-condensing); 70Kpa-106Kpa. |
Transport and Storage environment: | Temperature: -25°C ~ +70°C (-13°F ~ 158°F); Humidity: 15% ~ 93% RH (non-condensing). |
Thermometer set:
The unit must be utilized in accordance with current standards separately from domestic wastes. For utilizing it is necessary to contact special organizations licensed to make utilization.
Warranty period is 2 years from the date of purchase.
This warranty doesn't cover any damages caused by improper using, and also battery, protective cover and packaging. When a manufacturing defect is revealed during the warranty period a faulty unit would be repaired or, if repairing is impossible, replaced with another one.
Manufacturing date is on the unit in a serial number: last 2 fi gures of the year, then month number.
The manufacturer may change units partially or completely if necessary, without prior notice.
FOLLOW INSTRUCTIONS BEFORE USE
MANUFACTURER'S NAME
ARTICLE NUMBER
SERIAL NUMBER
TYPE BF EQUIPMENT
DISPOSAL FOR SEPARATE COLLECTION
CE MARK (0044)
OPERATING CONDITION, TEMPERATURE 5˚C ~ 40˚C
STORAGE CONDITION, TEPERATURE -25˚C ~ 70˚C
KEEP DRY
HOUSING INGRESS PROTECTION RATE: IP20
ELECTROMAGNETIC COMPATIBILITY INFORMATION
Manufacturer's Declaration of the Electromagnetic Compatibility
Table 1
Guidance and manufacturer's declaration – еlectromagnetic emissions
Emissions test | Compliance |
RF emissions CISPR 11 | Group 1 |
RF emissions CISPR 11 | Class B |
Harmonic emissions IEC 61000-3-2 | Not application |
Voltage fluctuations/ flicker emissions IEC 61000-3-3 | Not application |
Table 2
Guidance and manufacturer's declaration – еlectromagnetic immunity
Immunity test | IEC 60601-1-2 Test level | Compliance level |
Electrostatic discharge (ESD) IEC 61000-4-2 | ±8 kV contact ±2 kV, ±4 kV, ±8 kV, ±15 kV air | ±8 kV contact ±2 kV, ±4 kV, ±8 kV, ±15 kV air |
Electrical fast transient/ burst IEC 61000-4-4 | Power supply lines: ±2 kV input/output lines: ±1 kV | Not application |
Surge IEC 61000-4-5 | line(s) to line(s): ±1 kV. line(s) to earth: ±2 kV. 100 kHz repetition frequency | Not application |
Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11 | 0% 0.5 cycle At 0°, 45 °, 90 °, 135 °, 180 °, 225 °, 270 ° and 315 ° 0% 1 cycle And 70% 25/30 cycles Single phase: at 0 0% 300 cycle | Not application |
Power frequency magnetic fi eld IEC 61000-4-8 | 30 A/m 50Hz/60Hz | 30 A/m 50Hz/60Hz |
Conduced RF IEC 61000-4-6 | 150KHz to 80MHz: 3Vrms 6Vrms (in ISM and amateur radio bands) 80% Am at 1kHz | Not application |
Radiated RF IEC 61000-4-3 | 10 V/m 80 MHz – 2,7 GHz 80% AM at 1 kHz | 10 V/m 80 MHz – 2,7 GHz 80% AM at 1 kHz |
NOTE UT is the a.c. mians voltage prior to application of the test level. Remark In the presence of strong electromagnetic fi eld (Horizontal/Vertical, 80MHz~180MHz and 1.4GHz~2GHz, 10V/m applied on the product), it will have a change from 0.1°C to 0.3°C of the measured value. In case of malfunction due to these strong electromagnetic fi elds, user need to wait till resume normal operation without intervene. |
Table 3
Guidance and manufacturer's declaration – electromagnetic Immunity
Radiated RF IEC61000-4-3 (Test specifi cations for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment) | Test Frequency (MHz) | Band (MHz) | Service | Modulation | Modulation (W) | Distance (m) | IMMUNITY TEST LEVEL (V/m) |
385 | 380 – 390 | TETRA 400 | Pulse modulation 18 Hz | 1,8 | 0.3 | 27 | |
450 | 380 – 390 | GMRS 460, FRS 460 | FM ± 5 kHz deviation 1 kHz sine | 2 | 0.3 | 28 | |
710 | 704 – 787 | LTE Band 13, 17 | Pulse modulation 217 Hz | 0,2 | 0.3 | 9 | |
745 | |||||||
780 | |||||||
810 | 800 – 960 | GSM 800/900, TETRA 800, iDEN 820, CDMA 850, LTE Band 5 | Pulse modulation 18 Hz | 2 | 0.3 | 28 | |
870 | |||||||
930 | |||||||
1720 | 1 700 – 1 990 | GSM 1800; CDMA 1900; GSM 1900; DECT; LTE Band 1, 3, 4, 25; UMTS | Pulse modulation 217 Hz | 2 | 0.3 | 28 | |
1845 | |||||||
1970 |
Last revision 2020-W51
B.Well Swiss AG
Bahnhofstr. 24, 9443 Widnau, Switzerland
www.bwell-swiss.ch
Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.
Advertisement
Thank you! Your question has been received!
Need Assistance?
Do you have a question about the MED-3000 that isn't answered in the manual? Leave your question here.